0 |
EPROM Checksum: Replace Transmitter
or
Replace Transmitter |
1 |
RAM Error: Replace Transmitter |
2 |
Debug, IRQ: Inform Dräger
Fax: +49-451-882-2042 |
3 |
Panic: Unknown Bit Cmd 48, Byte 0, 0x10 |
4 |
Panic: Unknown Bit Cmd 48, Byte 0, 0x08 |
5 |
Debug, com1: Inform Dräger
Fax: +49-451-882-2042 |
6 |
Debug, com2: Inform Dräger
Fax: +49-451-882-2042 |
7 |
Panic: Unknown Bit Cmd 48, Byte 0, 0x01 |
8 |
No sensorConnect sensor
or
Check sensor for electrolyte leakage: Resistance between pins 3<->6 shall be > 10MOhms
or
Replace Sensor
or
Replace Transmitter |
9 |
Invalid sensor typeUse only sensors intended for Polytron2 |
10 |
Invalid sensor versionUse only sensors intended for Polytron2
or
Replace Transmitter |
11 |
Wrong sensor installedInstall Senser with correct Part No. or disable Sensor Lock temporarily |
12 |
Wrong Gas Database |
13 |
Invalid EEPROM versionPerform Standard Initialization
or
Perform Fundamental Initialization
or
Calibrate analog interface
or
Replace Transmitter |
14 |
Panic: Unknown Bit Cmd 48, Byte 1, 0x02 |
15 |
Panic: Unknown Bit Cmd 48, Byte 1, 0x01 |
16 |
I2C-bus error during sensor data uploadReplace Sensor
or
Replace Transmitter |
17 |
No valid sensor data read blockReplace Sensor
or
Replace Transmitter |
18 |
No valid sensor data notice blockReplace Sensor
or
Replace Transmitter |
19 |
No valid sensor data write blockReplace Sensor
or
Replace Transmitter |
20 |
Checksum of gas name table corruptedInterrupt supply voltage > 1 sec (Reset) |
21 |
Panic: Unknown Bit Cmd 48, Byte 2, 0x04 |
22 |
Panic: Unknown Bit Cmd 48, Byte 2, 0x02 |
23 |
Panic: Unknown Bit Cmd 48, Byte 2, 0x01 |
24 |
Manufacturer database corruptedReplace Transmitter |
25 |
EEPROM database corruptedPerform Fundamental Initialization
or
Calibrate analog interface |
26 |
I2C-bus error during sensor data downloadInterrupt supply voltage > 1 sec (Reset) |
27 |
Sensor stack 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
28 |
Sensor stack 2 ErrorInform Dräger
Fax: +49-451-882-2042 |
29 |
HC11 Stack 1: Inform Dräger
Fax: +49-451-882-2042 |
30 |
HC11 Stack 2: Inform Dräger
Fax: +49-451-882-2042 |
31 |
Panic: Unknown Bit Cmd 48, Byte 3, 0x01 |
32 |
Panic: Unknown Bit Cmd 48, Byte 4, 0x80 |
33 |
Panic: Unknown Bit Cmd 48, Byte 4, 0x40 |
34 |
Panic: Unknown Bit Cmd 48, Byte 4, 0x20 |
35 |
Analog Database Checksum: Interrupt supply voltage > 1 sec (Reset) |
36 |
Install test failedCheck supply voltage |
37 |
Warning Parameters corrupted: Inform Dräger
Fax: +49-451-882-2042 |
38 |
Invalid Supply Voltage: Check supply voltage |
39 |
Invalid Loop Current: Check supply voltage; Calibrate analog interface
or
Replace Transmitter |
40 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x80 |
41 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x40 |
42 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x20 |
43 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x10 |
44 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x08 |
45 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x04 |
46 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x02 |
47 |
Panic: Unknown Bit Cmd 48, Byte 5, 0x01 |
48 |
No valid span calibrationCalibrate Sensor (zero) and Calibrate Sensor (span) |
49 |
No valid zero calibrationCalibrate Sensor (zero) |
50 |
No write access to sensor database during calibrationInterrupt supply voltage > 1 sec (Reset) and Calibrate Sensor (zero) and Calibrate Sensor (span)
or
Replace Sensor
or
Replace Transmitter |
51 |
Panic: Unknown Bit Cmd 48, Byte 14, 0x10 |
52 |
Too negative concentrationClean ambient from gasses with negative cross sensitivity
or
Calibrate Sensor (zero) |
53 |
Sensor failed self testReplace Sensor |
54 |
Sensor potential incompatibleUse only sensors intended for Polytron2 |
55 |
Panic: Unknown Bit Cmd 48, Byte 14, 0x01 |
56 |
Debug measurement 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
57 |
Debug ADC 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
58 |
Debug ADC 2 ErrorInform Dräger
Fax: +49-451-882-2042 |
59 |
Incorrect sensor potentialCheck supply voltage
or
Replace Transmitter |
60 |
Pre-Amplifier Offset ErrorReplace Transmitter |
61 |
Pre-Amplifier power consumption ErrorReplace Transmitter |
62 |
Debug measurement 3 ErrorInform Dräger
Fax: +49-451-882-2042 |
63 |
ADC overflowReplace Transmitter
or
Interrupt supply voltage > 1 sec (Reset) |
64 |
ADC timing ErrorInform Dräger
Fax: +49-451-882-2042 |
65 |
Panic: Unknown Bit Cmd 48, Byte 16, 0x20 |
66 |
ADC common mode ErrorCheck supply voltage
or
Replace Transmitter |
67 |
ADC span ErrorCheck supply voltage
or
Replace Transmitter |
68 |
ADC zero ErrorCheck supply voltage
or
Replace Transmitter |
69 |
HC11 ADC span ErrorCheck supply voltage
or
Replace Transmitter |
70 |
HC11 ADC zero ErrorCheck supply voltage
or
Replace Transmitter |
71 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x80 |
72 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x40 |
73 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x20 |
74 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x10 |
75 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x08 |
76 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x04 |
77 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x02 |
78 |
Panic: Unknown Bit Cmd 48, Byte 17, 0x01 |
79 |
Panic: Unknown Bit Cmd 48, Byte 18, 0x80 |
80 |
Panic: Unknown Bit Cmd 48, Byte 18, 0x40 |
81 |
Panic: Unknown Bit Cmd 48, Byte 18, 0x20 |
82 |
Panic: Unknown Bit Cmd 48, Byte 18, 0x10 |
83 |
Panic: Unknown Bit Cmd 48, Byte 18, 0x08 |
84 |
ADC invalid paramtersInform Dräger
Fax: +49-451-882-2042 |
85 |
ADC time outInform Dräger
Fax: +49-451-882-2042 |
86 |
ADC overrunInform Dräger
Fax: +49-451-882-2042 |
87 |
Analog span not calibrated: Calibrate analog interface |
88 |
Analog zero not calibrated: Calibrate analog interface |
89 |
Panic: Unknown Bit Cmd 48, Byte 22, 0x20 |
90 |
Panic: Unknown Bit Cmd 48, Byte 22, 0x10 |
91 |
Panic: Unknown Bit Cmd 48, Byte 22, 0x08 |
92 |
Date not programmedSet Clock |
93 |
Time not programmedSet Clock |
94 |
Clock back-up supply downSet Jumper X3 or replace Li-battery
or
Replace Transmitter |
95 |
Panic: Unknown Bit Cmd 48, Byte 23, 0x80 |
96 |
Panic: Unknown Bit Cmd 48, Byte 23, 0x40 |
97 |
Sensor warm-up, no measurement: Wait for sensor to warm up |
98 |
Sensor warm-up, Measurement error <4x spec.: Wait for sensor to warm up |
99 |
Sensor counter electrode voltage out of rangeStop non permissable permanent gassing
or
Replace Sensor
or
Replace Transmitter
or
Replace Transmitter |
100 |
Gas concentration is too highReduce gas concentration
or
Replace Sensor |
101 |
Sensor temperature too lowAdjust ambient temperature to permissable range
or
Replace Sensor
or
Replace Transmitter |
102 |
Sensor temperature too highAdjust ambient temperature to permissable range
or
Replace Sensor
or
Replace Transmitter |
103 |
End of sensor life due to exposureReplace Sensor |
104 |
End of sensor life due to poor sensitivityReplace Sensor |
105 |
Panic: Unknown Bit Cmd 48, Byte 24, 0x20 |
106 |
Panic: Unknown Bit Cmd 48, Byte 24, 0x10 |
107 |
Too high exposure since last (span) calibrationCalibrate Sensor (zero) and Calibrate Sensor (span) |
108 |
Calibration interval expiredCalibrate Sensor (zero) and Calibrate Sensor (span) |
109 |
Gas concentration was too highCalibrate Sensor (zero) and Calibrate Sensor (span) |
110 |
Sensor temperature was too highCalibrate Sensor (zero) and Calibrate Sensor (span) |
111 |
Upload Var List |
112 |
4d |
113 |
2d |
114 |
4.2f |
115 |
8.2f |
116 |
.2f |
117 |
.1f |
118 |
7d |
119 |
.3f |
120 |
8.1f |
121 |
8.3f |
122 |
02X |
123 |
d |
124 |
3d |
125 |
5.2f |
126 |
2x |
127 |
p2GasIndex |
128 |
p2UnitIndex |
129 |
8u |
130 |
Panic: Unknown Bit Cmd 48, Byte 19, 0x10 |
131 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x10 |
132 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x08 |
133 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x04 |
134 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x02 |
135 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x01 |
136 |
EnterParamMode |
137 |
ExitParamMode |
138 |
EnterGasConfig |
139 |
ExitGasConfig |
140 |
Unexpected Error Command %{cmd} |
141 |
The handheld will temporarily loose communication with the Polytron. Please Select "Online" to Reconnect |
142 |
Online Info |
143 |
Mode %{1} %{2} %{3}
Online Info
%{4} %{5} %{6} %{7} %{8} %{14}
%{9} %{10} %{11} %{12} %{13} %{15} %{16} |
144 |
Master Reset |
145 |
Master Reset - Performs a master reset of the field device. Equivalent to cycling the transmitter power supply |
146 |
About to Perform Master Reset |
147 |
The handheld will temporarily loose communication with the Polytron. Please Select "Online" to Reconnect |
148 |
Master Reset in
progress please wait |
149 |
Device maybe stuck in Master Reset |
150 |
Master reset OK |
151 |
MaintPasswd=%{0}
Change Password? |
152 |
Enter Maintenance Password |
153 |
Entering Parameter Configuration Mode |
154 |
Saving Password |
155 |
Entering Measurement Mode |
156 |
ConfigPasswd=%{0}
Change Password? |
157 |
Enter Configuration Password |
158 |
Disable Warning Signal? |
159 |
Disabling Analog Warning Signal |
160 |
Enable Warning Signal? |
161 |
N/A! Multidrop |
162 |
Device is in Mutidrop; Analog Warning Signal is NOT Applicable |
163 |
Device is in Mutidrop; Analog Warning Signal is NOT Applicable |
164 |
N/A! Multidrop |
165 |
Current Loop about to be manipulated. Disable Alarms |
166 |
Set Analog Out |
167 |
4mA;20mA;Other mA |
168 |
Output Set to:
%{2} %{1} |
169 |
Use Special Signal |
170 |
Fault;Warning;Maintenance |
171 |
Output Set to
Fault Signal |
172 |
Output Set to
Warning Signal |
173 |
Output Set to
Maintenance Signal |
174 |
Enable Alarms |
175 |
About to Restore Factory Settings |
176 |
Restoration in
progress,
please wait |
177 |
Device maybe stuck Restoring |
178 |
Error restoring DB1 |
179 |
Error restoring DB 2 |
180 |
Error restoring DB 3 |
181 |
Restored Settings |
182 |
TestGas= %{0}
MaxConc= %{3} %{1} Enter Actual Conc |
183 |
Value Stabilized?
%{2} %0 %{1} |
184 |
Value OK?
%{2} %{0} %{1} |
185 |
Yes;No |
186 |
Not doing the calibration |
187 |
Too many write attempts |
188 |
Remove Test Gas |
189 |
Is Value out of Alarm Range?
%{2} %0 %{1} |
190 |
WARNING!
FACTORY USE ONLY!
For restoring a corrupted DB ONLY. Erases current loop calibration! |
191 |
The handheld will temporarily loose communication with the Polytron. Please Select "Online" to Reconnect |
192 |
WARNING!
FACTORY USE ONLY!
Initializes DB (after a SW-Update). |
193 |
History Buffer
%{4} %{5} %{6} %{7} %{8} %{14}
%{9} %{10} %{11} %{12} %{13} %{15} %{16}
Reset History? |
194 |
MeasGas= %{0}
Units= %{1}
Change Gas/Units? |
195 |
ExitGasSelect |
196 |
Exiting from Gas Selection (selection mode not active) |
197 |
Gas/Units NOT changed |
198 |
TestGas= %{0}
Units= %{1}
Change Test Gas or Units? |
199 |
ExitCalGasSelect |
200 |
Exiting from Cal Gas Selection (selection mode not active) |
201 |
Cal Gas and Units NOT changed |
202 |
ExitReloadDB |
203 |
Exiting from Cal Gas Selection (Reload DB) |
204 |
Reloading Measurement Gas DB |
205 |
Upload Timeout.
Continue Upload? |
206 |
Bad DB Upload.
Retry Upload? |
207 |
Offline |
208 |
Sensor ready in
%{0} Minutes
Calibration possible in
%{1} Minutes |
209 |
Warning: SEND will fail while HMI Active. Please Correct |