0 |
Sensor Serial No. |
1 |
7d |
2 |
Acknowledge Error |
3 |
Password |
4 |
Password
Enter a password to protect the detector against unauthorized access. Then access to editable parameters is disabled. To undo the protection again, you must enter the password again. You can choose any password you want; it may comprise a maximum of 8 characters or digits. |
5 |
Signal Unlocked |
6 |
Signal Unlocked
You will get alerted when the device is unlocked with a password. Choose WARNING only if the control system is to be alerted when the device was unlocked with the password.
OFF
The function is switched off.
WARNING
Once the detector is unlocked with a password, a warning is issued. If "Error Handling" is set to SENSITIVE, then a fault current is output in addition to the alarm. |
7 |
OFF |
8 |
No signaling when device is unlocked |
9 |
WARNING |
10 |
As long as device is unlocked it signals error current and error relais alarm. |
11 |
License Key Part1 |
12 |
License Key Part2 |
13 |
License Key Part3 |
14 |
License Key Part4 |
15 |
License Key Part5 |
16 |
Date |
17 |
Date
Enter here the current date.
The date is necessary for the proper work of the decay compensation and also for the error log and modification log, to know at which day the error or a parameter modification happened. |
18 |
Time |
19 |
Time
Enter here the current time.
The time is necessary for the error log and modification log, to know at which time the error or a parameter modification happened. |
20 |
Operating Mode |
21 |
Operating Mode
RUN
The measurement is in the normal measurement mode.
WARNING
A warning message has appeared. The error code is indicated in "Sensor Settings".
ERROR
An error is indicated. The error code is displayed in "Sensor Settings".
SHUTDOWN
A serious error is indicated. The error code is displayed in "Sensor Settings".
HOLD
The measurement is in the hold state, which means that the measured value and the current output signal is frozen. This status can occur under the following conditions:
- the plateau recording is running
- interference radiation was detected
- the digital input is on Hold
TEST
A test value is enabled. |
22 |
INIT |
23 |
INIT
Boot process. |
24 |
STARTUP |
25 |
STARTUP
Boot process. |
26 |
SHUTDOWN |
27 |
SHUTDOWN
A serious error is indicated. The error code is indicated in "Sensor Settings". |
28 |
ERROR |
29 |
ERROR
An error is indicated. The error code is indicated in "Sensor Settings". |
30 |
WARNING
A warning message has appeared. The error code is indicated in "Sensor Settings". |
31 |
HOLD |
32 |
HOLD
The measurement is in the hold state, which means that the measured value and the current output signal is frozen. This status can occur under the following conditions:
- the plateau recording is running
- interference radiation was detected
- the digital input is on Hold |
33 |
TEST |
34 |
TEST
A test value is enabled. |
35 |
RUN |
36 |
RUN
The measurement is in the normal measurement mode. |
37 |
Safety Mode |
38 |
Safety Mode
Indicates whether the safety mode is enabled:
ON
Safety mode is enabled.
OFF
Safety mode is disabled. |
39 |
undefined help string |
40 |
ON |
41 |
Error Handling |
42 |
Error Handling
Here you can set a different weighting of errors and error handling:
SENSITIVE
All faults cause the current output to report a Fault current. To also get warning messages, you must also evaluate the messages via the HART® signal or the digital output.
NORMAL
Only fatal errors are reported as a fault current. Thus, the measured via the current signal will fail only if the measurement can no longer be used. To also get minor error and warning messages, you must also evaluate the messages via the HART® signal or the digital output. |
43 |
SENSITIVE |
44 |
Measurement is aborted in case of error. |
45 |
NORMAL |
46 |
Measurement continues in case of error. |
47 |
Reset Device |
48 |
INACTIVE |
49 |
Reset is inactive. |
50 |
SW RESET |
51 |
Reset software now. |
52 |
FACTORY RESET |
53 |
Reset to factory settings now. |
54 |
MODIFICATION LOG RESET |
55 |
Clear modification log now. |
56 |
ERROR LOG RESET |
57 |
Clear error log now. |
58 |
Mod Log Index |
59 |
Mod Log Index
Active Index of modification log entry |
60 |
Description |
61 |
Description
Description of displayed modififcation log entry |
62 |
New Value |
63 |
New Value
The modified parameter has changed to this value |
64 |
Old Value |
65 |
Old Value
This was the former value of the modified parameter. |
66 |
Mod Date |
67 |
Mod Date
The date of mofification |
68 |
Modification Time |
69 |
Modification Time
The time of mofification |
70 |
Error Code |
71 |
Error Code
Indicates if an error is present in the active detector. If no error is present, then 0 is displayed here. An error message is displayed with a three digit number. The cause of the error and suggestions for correcting the error are described in the operating manual. |
72 |
Error Log Index |
73 |
Error Log Index
Active index of error log entry |
74 |
Error Date |
75 |
Error Date
Occurence date of active displayed log entry error |
76 |
Error Time |
77 |
Error Time
Occurence time of active displayed log entry error |
78 |
Error Description |
79 |
Error Description
Description of active displayed log entry error |
80 |
Bus Address Source |
81 |
AUTO DETECTION |
82 |
|
83 |
MENU |
84 |
Bus Address |
85 |
Bus Address
Enter fieldbus address of device. |
86 |
Device Temp. |
87 |
.1f |
88 |
Temp. Unit |
89 |
Device Temp. Test |
90 |
Device Temp. Test
Enter test value for device temperature |
91 |
Device Temp. Mode |
92 |
Device Temp. Mode
Select device temperature mode. Select 'Fix Value' for test. |
93 |
Temp. Upper Limit |
94 |
Temp. Lower Limit |
95 |
Temp. Hysteresis |
96 |
Device Temp. Min |
97 |
Device Temp. Min
Minimum device temperature since last reset |
98 |
Device Temp. Max |
99 |
Device Temp. Max
Maximum device temperature since last reset |
100 |
Reset Temp Extrema |
101 |
RESET |
102 |
Reset Temperature Extrema now. |
103 |
Detector Code |
104 |
Detector Code
The currently selected detector code is displayed. The detector code is crucial for the proper function of the automatic HV control (high voltage control). The detector code has already been set by BERTHOLD TECHNOLOGIES. It is dependent on the scintillator size and the detector type, see operating manual. |
105 |
HV Manual |
106 |
HV Manual
You can enter a fixed HV value. The value becomes active when you select MANUAL at "HV-Mode". |
107 |
.0f |
108 |
HV Default |
109 |
HV Default
This setting affects:
- A quick operation point control after power failure
After a power failure, the HV starts at the last "HV Average" value which results from the "HV Default". This will significantly reduce the start-up phase after power failure.
- Error detection in case of HV-drift
If the HV deviates too far (+40%, -20%) from the HV Default or if the HV reaches the limits of the range of 300 V or 1300 V, this is reported as an error.
- Special feature: 0V
For maintenance purposes, the HV Default can also be set to 0V. In this case, the mechanisms described above are suspended; after a power failure, the measurement begins to get adjusted at 800 V. "HV Default" has already determined and set by BERTHOLD TECHNOLOGIES. |
110 |
HV Live |
111 |
HV Live
Displays the current HV (high voltage) at the photomultiplier. If the "HV Mode" is set to AUTO (normal operation), the values in HV Live must change, viewed over several seconds. |
112 |
HV Feedback |
113 |
HV Average |
114 |
HV Average
Shows the average HV (high voltage) of the last 10 days. Significant deviations from "HV Live" to "HV Average" are considered and reported as errors. |
115 |
Cps Single Detector |
116 |
Cps Single Detector
Shows the count rate of the active detector normalized to one second. |
117 |
cps |
118 |
Meas CH CPS |
119 |
Control CH CPS |
120 |
.2f |
121 |
Auxiliary CH CPS |
122 |
HV Mode |
123 |
HV Mode
The HV-mode (high-voltage control) provides the following options:
AUTO
Automatic HV control. Select this mode in normal measurement mode. It enables a temperature-stable operation of the detector.
MANUAL
Manual HV control. This setting is used for testing purposes; you can also provisionally set the operating point of the detector. The specified voltage must, however, lie in the plateau. If you select "Manual", the automatic HV control is switched off. The HV is then set to the value that was set in the menu item HV Manual.
PLATEAU
Starts the plateau measurement. If you select PLATEAU, the measurement mode is exited and the plateau of the multiplier measured, which is used in the detector. |
124 |
AUTO |
125 |
Sensor works with HV auto mode |
126 |
MANUAL |
127 |
Sensor works with HV manual mode |
128 |
PLATEAU |
129 |
Sensor is gathering a plateau |
130 |
HV Start |
131 |
HV Start
Enter HV start value for plateau measurement beginning |
132 |
4u |
133 |
HV Stop |
134 |
HV Stop
Enter HV stop value for plateau measurement end |
135 |
HV Step |
136 |
HV Step
Enter HV step width for plateau measurement step width |
137 |
Meas. Time |
138 |
Meas. Time
Enter measuring time for each plateau measuring point |
139 |
Plateau Tab Entries |
140 |
Plateau Date |
141 |
Plateau Index |
142 |
Plateau Index
Active plateau table Index |
143 |
Plateau HV |
144 |
Plateau HV
HV of active displayed plateau table entry |
145 |
Plateau Impulses |
146 |
Plateau Impulses
Impulses of active displayed plateau table entry |
147 |
PMT Current |
148 |
PMT Current
Displays active current at multiplier |
149 |
.4f |
150 |
PMT Current Max |
151 |
PMT Current Max
Displays the maximum PMT current measured since last reset. |
152 |
Reset Current Extrema |
153 |
Reset PMT now. |
154 |
Error Current Value |
155 |
Error Current Value
Here you define the fault current in mA, if you have selected Value in the "Loop Alarm Type" parameter. If you have selected High or Low, then the appropriate current value is displayed here. If you exceed the mA value, then Value is set in the "Loop Alarm Type" parameter. |
156 |
Current Lower Limit |
157 |
Current Lower Limit
Lower limit of the current range for the 4-20 mA current output. For adjustment purposes or to safely identify an over or under measuring range condition, the current range available for the measurement signal is extended beyond the standard range of 4 mA and 20mA. According to the Namur specifications (NE 43), the lower current value must not be less than 3.8 mA. The limits can be set within the range from 3.8 mA to 4 mA. |
158 |
Current Upper Limit |
159 |
Current Upper Limit
Upper limit of the current range for the 4-20 mA current output. For adjustment purposes or to safely identify an over or under measuring range condition, the current range available for the measurement signal is extended beyond the standard range of 4 mA and 20mA. According to Namur specifications (NE 43), the upper current value must not be higher than 20.5mA. The limits can be set within the range from 20 mA to 20.5 mA. |
160 |
Digital Out State |
161 |
ALARM |
162 |
Digital Out Mode |
163 |
Digital Out Test State |
164 |
OPEN |
165 |
CLOSED |
166 |
Digital In State |
167 |
Digital In Mode |
168 |
Digital In Test State |
169 |
Meas Delay Time |
170 |
Meas Delay Time
If interference radiation detection is enabled, the measured values are output with a delay. The delay is adjustable from 0 to 5 seconds. The delay is required so that at the moment of detection the measured value remains unaffected, because in this case, increased count rates are already present which may distort the measured value. |
171 |
Io Factor |
172 |
Io Factor
Defines the threshold where the interference radiation detection responds. With the default value of 1.5, the threshold is 1.5 times the active empty count rate and is suitable for most applications. You can adjust the factor for the threshold yourself. Note that a value below 1.5 may trigger faulty switching caused by fluctuations in the measured values. Increasing the factor decreases the sensitivity. Increase the factor if false alarms are triggered by significant fluctuations in the measured values, for example, by agitator blades. |
173 |
Waiting Time |
174 |
Waiting Time
If interference radiation is detected, the measurement is "frozen" and will be released at the earliest after the waiting period is over. You have to enter the value for the waiting time in seconds. |
175 |
Cps Average Mode |
176 |
Cps Average Mode
Select cps average mode. Select 'Fix Value' for test. |
177 |
Cps Average Test |
178 |
Cps Average Test
Enter test value for Cps Average |
179 |
Cps Average |
180 |
Cps Average
The ReadIn rate is newly averaged in every measurement cycle. Standardized to one second. |
181 |
Cal/Comp Date |
182 |
Cal/Comp Date
Date of last calibration or compensation |
183 |
Restore Date |
184 |
Restore Date
Contains the validity date that is used for the function "Restore Upload". It can be loaded with the calibration data set or entered by hand. |
185 |
Nuclide |
186 |
Nuclide
The indication must be identical with the isotope (nuclide) that is used in your source. |
187 |
Cs-137 |
188 |
Decay compensation will be done with Cs-137. |
189 |
Co-60 |
190 |
Decay compensation will be done with Co-60. |
191 |
USER DEFINED |
192 |
Background |
193 |
Background
The term background refers to the background radiation which is present in the environment. The background radiation has to be measured so that the detector can correctly compensate for the decay of the radiation source. |
194 |
Background
The term background refers to the background radiation which is present in the environment. The background radiation has to be measured so that the detector can correctly compensate for the decay of the radiation source. Select "ReadIn" to read in the count rate for the background radiation. If you already know the value, you can enter it here. |
195 |
Nuclide
Enter here the isotope (nuclide) that is used in your source:
- Co-60
- Cs-137
- USER DEFINED
This entry controls the automatic decay compensation. The isotope used is listed on the type plate of the shielding and on your delivery documents. The item "USER DEFINED" allows you to use any isotope you want. In this case, additional parameters are requested: Specify the half-life under "Half Life Time" and the absorption coefficient at "Absorption". |
196 |
Half Life Time |
197 |
Half Life Time
The isotope's half life period |
198 |
a |
199 |
Warning Date |
200 |
Warning Date
Enter the date when you want to get the Source Exchange warning (format: MM/DD/YYYY) |
201 |
Remaining Time |
202 |
ReadIn Time |
203 |
ReadIn Time
Here you can define the period of time over which the count rate is to be read-in for each calibration point. The statistical variation of the count rate is averaged over this time period. A longer read-in time results in a better mean value. A time period of 30 sec (default setting) is usually reasonable. |
204 |
Application Level |
205 |
Invalid |
206 |
Level 1 |
207 |
Level 2 |
208 |
Level 3 |
209 |
Level 4 |
210 |
Level 5 |
211 |
Level 6 |
212 |
Level 7 |
213 |
Level 8 |
214 |
SW Revision |
215 |
SW Revision Date |
216 |
Current Loop Monitoring |
217 |
Current Loop Monitoring
Enable or disable the monitoring of the 4-20mA current signal. Monitoring checks whether the set current is actually flowing in the current loop, and signals an error if any deviation is detected. You have the following setting options:
ENABLED
Monitoring is enabled (default setting). Unless there are compelling reasons, you should keep this setting. The setting is automatically enabled when the Safety Mode is selected.
DISABLED
Monitoring is disabled. The HART Communicator outputs a corresponding message. |
218 |
DISABLED |
219 |
ENABLED |
220 |
Cps Live |
221 |
Cps Live
Display of active unaveraged count rate. (Cps live) |
222 |
Level |
223 |
Level
Displays active level value. |
224 |
Level Unit |
225 |
Device variable family |
226 |
Level Classification |
227 |
Level Data Quality |
228 |
Level Limit Status |
229 |
Level Family Status |
230 |
Time Constant |
231 |
Time Constant
Shows the active time constant. The value 0.0 leaves the count rate unfiltered. |
232 |
Cps Max |
233 |
Cps Min |
234 |
Minimum Span |
235 |
Level Mode |
236 |
Update Period |
237 |
PV |
238 |
PV
Primary Variable |
239 |
Detector Device Id |
240 |
Alarm State |
241 |
Alarm State
Displays, if Limit Value is exceeded. |
242 |
Digital Out Function |
243 |
LS ALARM |
244 |
Level switch alarm set |
245 |
DET. TEMP |
246 |
Detector temperature out of limits |
247 |
Measurement in hold mode |
248 |
WARNING + ERROR |
249 |
Measurement is in error mode |
250 |
BUILD UP |
251 |
Build up alarm detected |
252 |
RAD. INTERF. |
253 |
Radiation interference detected |
254 |
Digital In Function |
255 |
Digital input disabled |
256 |
Hold Measurement when digital input closed |
257 |
EMPTY |
258 |
Calibrate empty point when digital input is closed |
259 |
FULL |
260 |
alibrate full point when digital input is closed |
261 |
ReadIn Mode |
262 |
ReadIn disabled |
263 |
BACKGROUND |
264 |
ReadIn background count rate |
265 |
LOWER-PT |
266 |
ReadIn count rate of lower point |
267 |
UPPER-PT |
268 |
ReadIn count rate of upper point. |
269 |
Calibration Mode |
270 |
RECALL |
271 |
RESTORE |
272 |
CALIBRATE |
273 |
CALIBRATE + PRESET |
274 |
Calibration Status |
275 |
0- OK |
276 |
0- OK
The calibration carried out is OK. |
277 |
1- ERROR BACKGROUND |
278 |
1- ERROR BACKGROUND
The count rate of the background radiation is higher than that for the calibration points for empty or full (Cal Point 0% or Cal Point 100%). |
279 |
2- ERROR CALIBRATION POINT |
280 |
2- ERROR CALIBRATION POINT
The empty count rate is lower than the full count rate (Lower Cal Pt-Empty < Upper Cal Pt-Full). |
281 |
3- ERROR THRESHOLD / TIME CONSTANT |
282 |
3- ERROR THRESHOLD / TIME CONSTANT
The switching threshold (Threshold) is too close to one of the calibration points for empty (Lower Cal Pt-Empty) or full (Upper Cal Pt-Full) or the time constant was selected too small for the manual calibration. The error also occurs when the time constant exceeds 999 s. |
283 |
4- ERROR THRESHOLD OUT OF RANGE |
284 |
4- ERROR THRESHOLD OUT OF RANGE
The switching threshold (Threshold) plus or minus the hysteresis exceeds the allowable range of 0% ... 100%. |
285 |
5- ERROR BU TC |
286 |
5- ERROR BU TC
The time constant for the build-up alarm is above 14,400 seconds. This is either because the difference between empty count rate and full count rate is too low (Lower Cal Pt-Empty ? Upper Cal Pt-Full) or the limit value is too close to one of the calibration points for empty (Lower Cal Pt-Empty) or full (Upper Cal Pt-Full). |
287 |
6- DATE ERROR |
288 |
6- DATE ERROR
A date is still set to the default value of 1.1.2000. At Calibrate, check the Date parameter. At Restore, check the date in Restore Date. |
289 |
7- CHECK ERROR |
290 |
7- CHECK ERROR
The status of the calibration parameters could not be fully verified. In this case, repeat the calibration. If the error occurs again, replace the detector, at least the detector electronics. The above described plausibility check cannot prevent that the user may read-in incorrect data. If in doubt, please consult a BERTHOLD TECHNOLOGIES service technician. |
291 |
Measurement Status |
292 |
Calibration Method |
293 |
Calibration Method
1-POINT LIN
Use the 1-point calibration only if full calibration is not possible. In this method, an empty adjustment (level below the monitoring level) is required. Using additional inputs such as product density and measuring path the count rate for the full calibration (level above the monitoring level) is then determined automatically.
2-POINT EXP
An empty calibration (level below the monitoring level) and a full calibration (level above the monitoring level) is required for a two-point calibration. |
294 |
1-POINT |
295 |
Calibration with one point |
296 |
2-POINT |
297 |
Calibration with two points |
298 |
Absorption |
299 |
Calibration Method
Shows the selected calibration method. |
300 |
Lower Point |
301 |
Lower Cal Pt -Empty |
302 |
Lower Cal Pt -Empty
Count rate with empty container or at a level below the switching threshold. Select "ReadIn" to read in the count rate. If you already know the value, you can enter it here. |
303 |
Op Lower Point |
304 |
Op Lower Point
If you use the gas density compensation, the count rate which would be obtained under operating conditions is displayed here. The value shown cannot be changed and depends on:
- Count rate in "Lower Cal Pt-Empty"
- Setup Gas Density
- Op Gas Density
- Measuring Path
- Nuclide |
305 |
Upper Point |
306 |
Upper Cal Pt -Full |
307 |
Upper Cal Pt -Full
Count rate with full container or at a level above the switching threshold. If you have selected in "Calibration Method" the setting 1-POINT, the value for "Upper Cal Pt-Full" is automatically calculated from the value of "Lower Cal Pt-Empty" and the parameters defined in the "Product Conditions menu". Select "ReadIn" to read-in the count rate, if you have chosen the setting "2-POINT" at Calibration Method. If you already know the value, you can enter it here. |
308 |
Selection |
309 |
Selection
OFF: No message is output.
DATE: The message Source Exchange is output on a specific date. Enter the date in "Warning Date".
TIME CONSTANT: In Max Time Constant enter a maximum time constant when a certain response time of the measurement should not be exceeded. (The response time is approximately equal to the time constant.) After entering the maximum time constant, the detector automatically calculates the date of the source exchange. The date is displayed under Date. To use the function TIME CONSTANT, you must set "Evaluation Mode" to AUTO SET. In the MANUAL setting the time constant is not adjusted automatically. |
310 |
Source exchange signaling is disabled |
311 |
DATE |
312 |
Source exchange depends on date |
313 |
TIME CONSTANT |
314 |
Source exchange depends on maximum time constant. |
315 |
Max Time Constant |
316 |
Max Time Constant
The maximum allowable time constant must be greater than the current, automatically calculated time constant (see "Time Constant"). The value of the time constant approximately corresponds to the response time of the measurement. After entry of a value, the date on which the warning will be expected is displayed in "Warning Date". Due to the automatic decay compensation, the time constant is increased constantly until the maximum time constant is reached, whereupon the source exchange warning message is reported. |
317 |
Response Mode |
318 |
Response Mode
Select the detection mode:
DISABLED
Interference radiation detection and Rapid switchover are disabled.
RAD. INTERFERENCE
Interference radiation detection is enabled. |
319 |
Radiation Interference is disabled |
320 |
RAD.INTERFERENCE |
321 |
Radiation Interference |
322 |
Evaluation Mode |
323 |
Evaluation Mode
Threshold and Hysteresis are calculated automatically and adjusted continuously, or whether they can be manually set to a fixed value.
AUTO SET
This is the default setting which should ideally be maintained. It ensures that the above parameters are optimized and adjusted continuously and checked at every decay compensation.
MANUAL SET
Use this setting only if it is absolutely necessary for your application and if you are aware of the consequences of manual input of fixed values. Incorrect entries that jeopardize the switching reliability can be compensated for only partially in this mode.
Important: With manual input (MANUAL SET), the decay compensation has to be taken into account. Therefore, you must choose the time constant so large that it is valid for years to come and will not lead to incorrect switching. In the default setting (AUTO SET), the switching reliability is checked as well in the daily decay compensation. |
324 |
AUTO SET |
325 |
The dynamic reaction will be done automatically. |
326 |
MANUAL SET |
327 |
Configure the dynamic reaction manually. |
328 |
Hysteresis |
329 |
Hysteresis
Shows the current hysteresis. The hysteresis is calculated automatically, but can also be manually set to a fixed value (Evaluation Mode: MANUAL SET). The hysteresis lies symmetrically around the switching threshold and prevents switching back and forth, which would be triggered by fluctuations of the measured value. |
330 |
Threshold |
331 |
Threshold
Displays the current switching threshold (alarm threshold for the level). The time constant is calculated automatically (factory setting), but can also be manually set to a fixed percentage value (Evaluation Mode: MANUAL SET). We recommend using the automatic calculation (AUTO SET), unless important reasons speak against it. If the alarm for the switching threshold is triggered, the following signals are set:
- the current output goes to 4 or 20 mA
- Alarm State is ON
- the open collector switches |
332 |
Threshold Cps |
333 |
Time Constant
Shows the current time constant. The time constant is either calculated automatically (factory setting), or manually set to a fixed value. The time constant smoothes the output signal. Statistical fluctuations, as well as level fluctuations due to the process, e.g. by agitators, are smoothed. The value of the time constant corresponds approximately to the response time of the measurement. Depending on the setting in "Evaluation Mode", either the calculated time constant is displayed (AUTO SET) or a time can be entered in seconds (MANUAL SET). We recommend using the automatic calculation (AUTO SET), unless important reasons speak against it. For manual input please keep in mind: The time constant must be sufficiently large to avoid incorrect switching. At the same time, however, an increase in the time constant will delay the reaction of the measurement to a level exceedance; therefore, the time constant should not be selected too large. A time constant of 20 seconds is a good choice for standard applications and correct source detector configurations. |
334 |
Switch Function |
335 |
Switch Function
You can specify whether the limit switch is used as a High Alarm or as a Low Alarm. The switching direction affects the fail-safe function of the digital output (open collector). To output the alarm as a signal via the open collector, you have to set "Digital Out Function" in the I/O Setup menu to LS ALARM. You have the following setting options:
MAX
The alarm is triggered when exceeding the switching threshold.
MIN
The alarm is triggered when falling below the switching threshold
MAX + BUILD UP
The alarm is triggered if the switching threshold is exceeded (High Alarm). In addition, monitoring is performed which triggers an alarm if wall deposits 1 is detected. Prerequisite for correct function is that both the count rate for empty and the count rate for full was measured. During the measurement of the empty value it must be ensured that no build-up occurs. |
336 |
MAX |
337 |
Application signals alarm, when measurement value exceeds threshold. |
338 |
MAX + BUILD UP |
339 |
Application signals alarm, when measurement value exceeds threshold and build up is detected. |
340 |
MIN |
341 |
Application signals alarm, when measurement value is below threshold. |
342 |
BU Alarm |
343 |
BU Alarm
Display build up alarm. |
344 |
Build up alarm is inactive. |
345 |
ACTIVE |
346 |
Build up alarm is active. |
347 |
BU Level |
348 |
BU Level
Second level measurement value parallel to normal level.(Averaged with 10x time constant) |
349 |
BU Threshold |
350 |
BU Threshold
Build up Threshold (+ Hysteresis) has to be exceeded for alarm signaling. |
351 |
BU Hysteresis |
352 |
BU Hysteresis
Needed for build up alarm signaling and to calculate BU Time Constant. |
353 |
BU Time Constant |
354 |
BU Time Constant
Equivalent to 10x time constant of level value. |
355 |
sec |
356 |
BU Delay Time |
357 |
Measuring Path |
358 |
Measuring Path
This menu item is required for "Calibration Method" 1-POINT to determine the full count rate automatically. The value to be entered is the length of the beam path through the measured product.
It is also required when using the function "Gas Density Compensation" and "Bulk Cone Measurement". |
359 |
Product Density |
360 |
Product Density
This setting is needed only when using "Calibration Method" 1-POINT or in connection with "Bulk Cone Diameter". For liquids you have to enter the liquid density in kg/m³, for bulk material, the bulk density. If you known the fluid density and the bulk density only approximately, then you have to enter the lowest expected value. |
361 |
Bulk Cone Diameter |
362 |
Bulk Cone Diameter
The setting is used to determine the switching point in bulk materials. In bulk materials the switching threshold should often lie at a specific bulk cone diameter. Enter a value greater than 0mm to calculate the switching threshold using the bulk cone diameter, provided that the "Measuring Path" and the "Product Density" are indicated. Please note that the time constant can be very large if the bulk cone diameter is selected too small or too large. |
363 |
Setup Gas Density |
364 |
Setup Gas Density
Enter the gas pressure in kg/m3 which is present during calibration. Use this parameter only if the gas density during calibration is different from the gas density under operating conditions. Using the gas density, a compensated count rate is then calculated from the count rate in "Lower Cal Pt-Empty" which you can read off under "Op Lower Point".
Prerequisite for the proper correction are correct entries in the following menu items:
- Setup Gas Density
- Op Gas Density
- Measuring Path
- Nuclide |
365 |
Gas Density at Op |
366 |
Gas Density at Op
Visible only if "Compensation Mode" is set to MANUAL GDA. Enter the gas pressure in kg/m3 you expect under operating conditions. Use this parameter only if the gas density during calibration is different from the gas density under operating conditions. Using the gas density, a compensated count rate is then calculated from the count rate in "Lower Cal Pt-Empty" which you can read off under "Op Lower Point". See manual for more details. |
367 |
Safety Status 1 |
368 |
Safety Status 1
Safety status flags part 1 |
369 |
Indicates ON when all following states are set to OK. |
370 |
Watchdog |
371 |
Indicates ON when the watchdog jumper on the CPU board is plugged. |
372 |
Err. Handling |
373 |
Indicates ON if the parameter Error Handling is set to SENSITIVE. |
374 |
Current Feedback |
375 |
Indicates ON if the monitoring of the Current Loop is activated. |
376 |
Loop Alarm = HIGH? |
377 |
Indicates ON if the parameter Loop Alarm Type is set to HIGH. |
378 |
Loop Test disabled? |
379 |
Indicates ON if the Loop Test is disabled. |
380 |
Multidrop disabled? |
381 |
Indicates ON if the Multidrop Mode is disabled. |
382 |
Safety Status 2 |
383 |
Safety Status 2
Safety status flags part 2 |
384 |
HV Mode AUTO? |
385 |
Indicates ON if the parameter HV Mode is set to AUTO. |
386 |
Indicates ON if HV Default is in the range between 300 to 1300V. |
387 |
Dig. In disabled? |
388 |
Indicates ON if the parameter Digital In Function is set to OFF. |
389 |
Dig. In Test disabled? |
390 |
Indicates ON if the parameter Digital In Mode in I/O Test Settings is set to NORMAL. |
391 |
Dig. Out Test disabled? |
392 |
Indicates ON if the parameter Digital Out Mode in I/O Test Settings is set to NORMAL. |
393 |
Safety Status 3 |
394 |
Safety Status 3
Safety status flags part 3 |
395 |
Cps Test disabled? |
396 |
Indicates ON if the parameter Cps Average Mode (test pulse generator) in Test Settings is set to NORMAL. |
397 |
Cps Comp Test disabled? |
398 |
Indicates ON if the parameter Cps Comp Average Mode (test pulse generator) in Test Settings is set to NORMAL. |
399 |
Dev Temp Test disabled? |
400 |
Indicates ON if the parameter Device Temp. Mode is set to NORMAL. |
401 |
Comp Mode disabled? |
402 |
Indicates ON if the parameter Compensation Mode in menu Cal Parameter/Product Conditions is set to OFF. |
403 |
Reset Modification Flag |
404 |
Software Reset |
405 |
Software Reset
This method performs a software reset on the active probe. |
406 |
Factory Reset |
407 |
Factory Reset
This method performs a factory reset on the active probe. |
408 |
Reset Error Log |
409 |
Reset Modification Log |
410 |
Safety OFF |
411 |
Safety OFF
Disables the safety mode. Keep the password handy in case you want to make changes, because you will need any time you make changes. If you disable the safety mode, the following settings are made automatically:
- the detector is unlocked, i.e. Write Protect is set to OFF
- Error Handling is set to NORMAL
- Response Mode is set to DISABLED |
412 |
Refresh Plateau |
413 |
Refresh Plateau
Read values from device and refresh graphical view of plateau table. |
414 |
Refresh Error Log |
415 |
Refresh Modification Log |
416 |
Set Poll Addr |
417 |
Set License Key |
418 |
Refresh Safety Statuses |
419 |
Safety ON |
420 |
Safety ON
The safety mode has to be enabled for safety-relevant applications. In safety mode, safety-limiting values are set to a safety-compatible setting. In addition, you will be prompted to protect the detector with a password. The following settings are set automatically when safety mode is enabled:
Current Loop Monitoring: = ENABLED
HV Mode = AUTO
Test Settings = NORMAL
Error Handling = SENSITIV
Digital inputs: = OFF
Multidrop: Poll Addr = 0 |
421 |
ReadIn |
422 |
ReadIn
Start reading-in the count rates. While the count rate is read in, the average is calculated and displayed continuously. Under "ReadIn Time" you define the read-in period. First choose which calibration point you want to read in:
BACKGROUND
The background radiation has to be measured so that the detector can correctly compensate for the decay of the radiation source. Before taking measurements, make sure you do not measure any radiation from the radiation source The easiest way to ensure this is if the source is not yet mounted. Otherwise, close the beam path and fill the container in addition.
LOWER-PT
For this measurement, the level in the tank must be below the monitoring level or the tank must be completely empty.
UPPER-PT
This measurement can only be performed at a 2-point calibration. For the measurement, the level in the tank must be above the monitoring level or the tank must be completely filled. |
423 |
Calibrate |
424 |
Calibrate
With this menu item you enable the calibration data determined during the measurements. The calibration data are transferred to "Meas Parameter". Thus, the detector will get a new calibration which in the future will be used to determine the measured values. After the calibration, a status message is displayed, indicating if the activation of the calibration data has been carried out successfully. If not, the measurement parameters are unchanged. |
425 |
Calibrate + Preset |
426 |
Calibrate + Preset
With this menu item you enable the calibration data determined during the measurements. In this case, the calibration data are transferred to "Meas Parameter". Thus, the detector will get a new calibration which in the future will be used to determine the measured values. After the calibration, a status message is displayed, indicating if the activation of the calibration data has been carried out successfully. If not, the measurement parameters are unchanged.
The following settings are made automatically at Calibrate + Auto Set:
Evaluation Mode = AUTO SET
OP Gas Density = 0
Setup Gas Density = 0
Bulk Cone = 0 |
427 |
Restore Upload |
428 |
Restore Upload
Has a similar function as "Calibrate", but considers a validity date, for example, of older calibration data and automatically performs a decay compensation. First, you have to save the original set of parameters in the HOST or the HART® Communicator as off-line parameters. The off-line parameters can then be restored to the detector. Then enable this parameter set with Restore Upload. The validity date (Restore Date) of the calibration, which is contained in the off-line parameters, will automatically be compared with the current date in the detector and the calibration count rate is adjusted accordingly. |
429 |
Recall |
430 |
Recall
Recall allows you to copy the current data set from Meas Parameter to Cal Parameter. This allows you to revise the valid parameters, without this having an influence on the measurement at first. After you have finished editing, you can activate the changed settings with "Calibrate". Recall is also necessary because the count rates in the calibration parameters (Cal Parameter) in contrast to the count rates of the measuring parameters (Meas Parameter) are not decay compensated. Important: "Recall" overwrites all settings of the "Cal Parameter" menu. |
431 |
Plateau Curve |
432 |
Plateau |
433 |
HV |
434 |
V |
435 |
Cps |
436 |
Plateau Table |
437 |
Point |
438 |
Error Log Table |
439 |
Modification Log Table |
440 |
Berthold Device |
441 |
% |
442 |
Current |
443 |
mA |
444 |
Temp. |
445 |
Trend |
446 |
Online - Root |
447 |
Device Config |
448 |
Setup |
449 |
Setup !locked! |
450 |
Setup locked, while write protection active |
451 |
Access |
452 |
Identification |
453 |
Location |
454 |
Device Information |
455 |
Device Revision |
456 |
Temp. Threshold Settings |
457 |
Current Output |
458 |
Digital Input |
459 |
HART Interface |
460 |
Service |
461 |
Sensor Temperature |
462 |
License Key |
463 |
Operating Status |
464 |
Device Variables Status |
465 |
Config Change Status |
466 |
Log |
467 |
Modification Log |
468 |
Test |
469 |
Error Status 100..207 |
470 |
Error Status 300..323 |
471 |
Error Status 324..507 |
472 |
Error Status 600..707 |
473 |
Error Status 900..907 |
474 |
Offline - Root |
475 |
Cal Parameter |
476 |
I/O Setup |
477 |
Live Display |
478 |
Diagnostic |
479 |
Safety |
480 |
Error Status |
481 |
Date - Time |
482 |
Sensor Configuration |
483 |
PMT |
484 |
Sensor Settings |
485 |
Adapt Calibration |
486 |
Cal Settings |
487 |
Cal Points |
488 |
Product Conditions |
489 |
Meas Parameter |
490 |
Meas Settings |
491 |
Meas Data |
492 |
Build Up Alarm |
493 |
Signal Condition |
494 |
Signal Parameter |
495 |
Signal Dependency |
496 |
Source Exchange |
497 |
Digital Output |
498 |
Plateau View |
499 |
Plateau Information |
500 |
Plateau Measurement |
501 |
Test Settings |
502 |
I/O Test Settings |
503 |
Error Log |
504 |
Quick Start |
505 |
Step 1 - Date |
506 |
Step 2 - Setup |
507 |
Step 3 - Product Conditions |
508 |
Step 4 - Calibration Points |
509 |
Parameter Setup |
510 |
Parameter Upload |
511 |
Maintenance |
512 |
Device Type |
513 |
LB480 LEVEL SWITCH |
514 |
Write Protect |
515 |
Write Protect
Indicates whether the detector is protected against changes in the settings (parameters).
NO
The detector is not write protected, so that the settings can be edited.
YES
The detector is write protected; settings cannot be edited, but they can still be displayed. |
516 |
Loop Current Mode |
517 |
FIXED VALUE |
518 |
Current Out |
519 |
8.2f |
520 |
Loop Alarm Type |
521 |
Loop Alarm Type
Here you specify the fault current, i.e. the current that is to be output in case of error.
Important: In Safety Mode, the fault current is automatically set to High.
You have the following setting options:
High
In case of error the current output is set to 22mA.
Low
In case of error the current output is set to 2mA.
Hold Last Value
In case of error the current output holds the last measured value.
Value
In case of error, the current output is set to the current value, which is set in Error Current Value. |
522 |
HIGH |
523 |
Current output in case of error: 22mA |
524 |
LOW |
525 |
Current output in case of error: 3.5 mA |
526 |
HOLD LAST VALUE |
527 |
Current output in case of error: Hold last value |
528 |
VALUE |
529 |
Current output in case of error: Error current value |
530 |
Transfer Function |
531 |
Transfer Function
Here you can specify how the current output should behave in case of a change in level:
LINEAR
A continuous 4-20 mA signal is output which is proportional to the count rate.
DISCRETE SWITCH
Depending on the set switching threshold (alarm threshold for the level) a current of 4 mA or 20 mA is output. Values between 4 and 20 mA will not be output. |
532 |
LINEAR |
533 |
DISCRETE SWITCH |
534 |
% Meas Range |
535 |
Upper Range Value |
536 |
Lower Range Value |
537 |
Tag |
538 |
Descriptor |
539 |
Message |
540 |
Device Id |
541 |
Poll Addr |
542 |
Final Assembly Num |
543 |
Field Dev Rev |
544 |
Software Rev |
545 |
Hardware Rev |
546 |
Num Req Preams |
547 |
Num Resp Preams |
548 |
Modification Counter |
549 |
Errors 101..108 |
550 |
Errors 101..108
System Errors Group 101..108 |
551 |
101 HW module missing or not tested |
552 |
Error No. 101 HW module missing or not tested |
553 |
102 Device data-set error |
554 |
Error No. 102 Device data-set error |
555 |
103 RAM Error |
556 |
Error No. 103 RAM Error |
557 |
104 Device Error |
558 |
Error No. 104 Device Error |
559 |
105 Real time clock not valid |
560 |
Error No. 105 Real time clock not valid |
561 |
106 Test mode active |
562 |
Error No. 106 Test mode active |
563 |
107 Watchdog reset |
564 |
Error No. 107 Watchdog reset |
565 |
108 Safety parameter invalid |
566 |
Error No. 108 Safety parameter invalid |
567 |
Errors 200..207 |
568 |
Errors 200..207
Mainboard Errors Group 200..207 |
569 |
200 Data flow |
570 |
Error No. 200 Data flow |
571 |
201 Error by analog input calibration |
572 |
Error No. 201 Error by analog input calibration |
573 |
202 Clock signal deviation |
574 |
Error No. 202 Clock signal deviation |
575 |
203 Reserved |
576 |
Error No. 203 Reserved |
577 |
204 Reserved |
578 |
Error No. 204 Reserved |
579 |
205 Reserved |
580 |
Error No. 205 Reserved |
581 |
206 Reserved |
582 |
Error No. 206 Reserved |
583 |
207 Reserved |
584 |
Error No. 207 Reserved |
585 |
Errors 300..307 |
586 |
Errors 300..307
Detector Control Errors Group 300..307 |
587 |
300 Data flow |
588 |
Error No. 300 Data flow |
589 |
301 Error by ADC calibration |
590 |
Error No. 301 Error by ADC calibration |
591 |
302 Error by DAC calibration |
592 |
Error No. 302 Error by DAC calibration |
593 |
303 Supply 5.0V |
594 |
Error No. 303 Supply 5.0V |
595 |
304 Reference 2.0V |
596 |
Error No. 304 Reference 2.0V |
597 |
305 Reference 2.5V |
598 |
Error No. 305 Reference 2.5V |
599 |
306 ERROR GND CPU ST9 |
600 |
Error No. ERROR GND CPU ST9 |
601 |
307 No impulses in measurement channel |
602 |
Error No. 307 No impulses in measurement channel |
603 |
Errors 308..315 |
604 |
Errors 308..315
Detector Control Errors Group 308..315 |
605 |
308 No impulses in control channel |
606 |
Error No. 308 No impulses in control channel |
607 |
309 No impulses in auxiliary channel |
608 |
Error No. 309 No impulses in auxiliary channel |
609 |
310 Impulse difference (measurement channel) |
610 |
Error No. 310 Impulse difference in measurement channel too large |
611 |
311 Impulse difference (control channel) |
612 |
Error No. 311 Impulse difference in control channel too large |
613 |
312 Impulse difference (auxiliary channel) |
614 |
Error No. 312 Impulse difference in auxiliary channel too large |
615 |
313 Instable pulse rates in both channels |
616 |
Error No. 313 Instable pulse rates in both channels |
617 |
314 Threshold of measurement channel 1 |
618 |
Error No. 314 Threshold of measurement channel 1 |
619 |
315 Threshold of measurement channel 2 |
620 |
Error No. 315 Threshold of measurement channel 2 |
621 |
Errors 316..323 |
622 |
Errors 316..323
Detector Control Errors Group 316..323 |
623 |
316 Threshold of control channel 1 |
624 |
Error No. 316 Threshold of control channel 1 |
625 |
317 Threshold of control channel 2 |
626 |
Error No. 317 Threshold of control channel 2 |
627 |
318 Threshold of auxiliary channel 1 |
628 |
Error No. 318 Threshold of auxiliary channel 1 |
629 |
319 Threshold of auxiliary channel 2 |
630 |
Error No. 319 Threshold of auxiliary channel 2 |
631 |
320 HV voltage |
632 |
Error No. 320 HV voltage |
633 |
321 Generated HV voltage |
634 |
Error No. 321 Generated HV voltage |
635 |
322 HV reached it's limit value |
636 |
Error No. 322 HV reached it's limit value |
637 |
323 HV average is 20% lower than default HV |
638 |
Error No. 323 HV average is 20% lower than default HV |
639 |
Errors 324..331 |
640 |
Errors 324..331
Detector Control Errors Group 324..331 |
641 |
324 HV average is 40% higher than default HV |
642 |
Error No. 324 HV average is 40% higher than default HV |
643 |
325 PMT current warning |
644 |
326 PMT current error |
645 |
327 Temperature sensors deviation |
646 |
Error No. 327 Temperature sensors deviation |
647 |
328 Temperature warning |
648 |
Error No. 328 Temperature warning |
649 |
329 Temperature out of allowed limits |
650 |
Error No. 329 Temperature out of allowed limits |
651 |
330 Detector malfunction |
652 |
Error No. 330 Detector malfunction |
653 |
331 Reserved |
654 |
Help 331 Reserved |
655 |
Errors 400..407 |
656 |
Errors 400..407
Interface Errors Group 400..407 |
657 |
400 Data flow |
658 |
Error No. 400 Data flow |
659 |
401 Supply 11V |
660 |
Error No. 401 Supply 11V |
661 |
402 Supply 5V |
662 |
Error No. 402 Supply 5V |
663 |
403 Supply 5VM |
664 |
Error No. 403 Supply 5VM |
665 |
404 Supply 3.3V |
666 |
Error No. 404 Supply 3.3V |
667 |
405 RS-485 Communication error |
668 |
Error No. 405 RS-485 Communication error |
669 |
406 Remote device warning |
670 |
Error No. 406 Remote device warning |
671 |
407 Remote device error |
672 |
Error No. 407 Remote device error |
673 |
Errors 500..507 |
674 |
Sensor Errors Group 500..507 |
675 |
500 Data flow |
676 |
Error No. 500 Data flow |
677 |
501 Undefined |
678 |
Help 501 Undefined |
679 |
502 Digital input malfunction |
680 |
Error No. 502 Digital input malfunction |
681 |
503 Digital output malfunction |
682 |
Error No. 503 Digital output malfunction |
683 |
504 Inherited Message |
684 |
Error No. 504 |
685 |
505 Undefined |
686 |
Help 505 Undefined |
687 |
506 Undefined |
688 |
Help 506 Undefined |
689 |
507 Reserved |
690 |
Help 507 Undefined |
691 |
Errors 600..607 |
692 |
Errors 600..607
Application Errors Group 600..607 |
693 |
600 Data flow |
694 |
Error No. 600 Data flow |
695 |
601 License Key Error |
696 |
Error No. 601 License Key Error |
697 |
602 Undefined |
698 |
Help 602 Undefined |
699 |
603 Meas Error, check <Error Status> |
700 |
Error No. 603 Meas Error, check <Error Status> |
701 |
604 Decay compensation error |
702 |
Error No. 604 Decay compensation error |
703 |
605 Source Exchange |
704 |
Error No. 605 Source Exchange |
705 |
606 Radiation interference |
706 |
Error No. 606 Radiation interference |
707 |
607 RID Interference |
708 |
Error No. 607 RID Interference |
709 |
Errors 608..615 |
710 |
Application Errors Group 608..615 |
711 |
608 Build up detected |
712 |
Error No. 608 Build up detected |
713 |
609 Inherited Message |
714 |
Error No. 609 |
715 |
610 Inherited Message |
716 |
Error No. 610 |
717 |
611 Inherited Message |
718 |
Error No. 611 |
719 |
612 Inherited Message |
720 |
Error No. 612 |
721 |
613 Inherited Message |
722 |
Error No. 613 |
723 |
614 Undefined |
724 |
Help 614 Undefined |
725 |
615 Undefined |
726 |
Help 615 Undefined |
727 |
Errors 700..707 |
728 |
Errors 700..707
Current Output Errors 700..707 |
729 |
700 Data flow |
730 |
Error No. 700 Data flow |
731 |
701 Current Loop Monitoring |
732 |
Error No. 701 Current Loop Feedback Difference |
733 |
702 Current loop malfunction |
734 |
Error No. 702 Current loop malfunction |
735 |
703 Current Loop Monitoring DISABLED |
736 |
Error No. 703 Current Loop Monitoring DISABLED |
737 |
704 ERROR GND CPU ST6 |
738 |
Error No. 704 ERROR GND CPU ST6 |
739 |
705 Reserved |
740 |
Help 705 Reserved |
741 |
706 Reserved |
742 |
Help 706 Reserved |
743 |
707 Reserved |
744 |
Help 707 Reserved |
745 |
Errors 900..907 |
746 |
Errors 900..907
HART Errors Group 900..907 |
747 |
900 Data flow |
748 |
Error No. 900 Data flow |
749 |
901 Signal Unlocked |
750 |
Error No. 901 Signal Unlocked |
751 |
902 Reserved |
752 |
Help 902 Reserved |
753 |
903 Reserved |
754 |
Help 903 Reserved |
755 |
904 Reserved |
756 |
Help 904 Reserved |
757 |
905 Reserved |
758 |
Help 905 Reserved |
759 |
906 Reserved |
760 |
Help 906 Reserved |
761 |
907 Reserved |
762 |
Help 907 Reserved |
763 |
Lock/Unlock Device |
764 |
Lock Dev Status |
765 |
Device is Locked |
766 |
Lock is Permanent |
767 |
Locked by Primary Master |
768 |
D/A trim |
769 |
D/A trim
Allows you to adjust the current output. You will need a current meter which has to be connected to the current loop. A method will take you through the adjustment. |
770 |
Loop test |
771 |
PV Analog Channel Saturated |
772 |
PV Analog Channel Fixed |
773 |
Wireless |
774 |
WirelessHART Device |
775 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
776 |
Primary |
777 |
Primary Analog Channel |
778 |
Secondary |
779 |
Secondary Analog Channel |
780 |
Tertiary |
781 |
Tertiary Analog Channel |
782 |
Quaternary |
783 |
Quaternary Analog Channel |
784 |
Quinary |
785 |
Quinary Analog Channel |
786 |
Critical Power Failure |
787 |
Lock all |
788 |
Trim not supported |
789 |
Configuration cannot be changed |
790 |
Locked by gateway |
791 |
Analog Input Channel |
792 |
Analog Output Channel |
793 |
Secondary Analog Channel Saturated |
794 |
Tertiary Analog Channel Saturated |
795 |
Quaternary Analog Channel Saturated |
796 |
Quinary Analog Channel Saturated |
797 |
Secondary Analog Channel Fixed |
798 |
Tertiary Analog Channel Fixed |
799 |
Quaternary Analog Channel Fixed |
800 |
Quinary Analog Channel Fixed |
801 |
Simulation active |
802 |
The device is in simulation mode and the level output is not representative of the process. |
803 |
Non-Volatile memory failure |
804 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
805 |
Volatile memory error |
806 |
The RAM memory check is invalid or maybe corrupt |
807 |
Watchdog reset executed |
808 |
A watchdog reset has been performed |
809 |
Voltage conditions out of range |
810 |
A voltage condition is outside its allowable range |
811 |
Environmental conditions out of range |
812 |
An internal or environmental condition is beyond acceptable limits. |
813 |
Electronic failure |
814 |
A hardware problem not related to the sensor has been detected. |
815 |
Subdevice list changed |
816 |
Duplicate master detected |
817 |
Capacity Denied |
818 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
819 |
Bandwidth allocation pending |
820 |
Block transfer pending |
821 |
Battery life |
822 |
Percent range |
823 |
Loop current |
824 |
Primary variable |
825 |
Secondary variable |
826 |
Tertiary variable |
827 |
Quaternary variable |
828 |
Read receive time |
829 |
Write date and time |
830 |
Non-Volatile Clock |
831 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
832 |
Clock Uninitialized |
833 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
834 |
None |
835 |
No restrictions |
836 |
SI Only |
837 |
Restricted to SI Unit codes only |
838 |
Process automation device |
839 |
Discrete device |
840 |
Hybrid:Process automation+discrete |
841 |
I/O System |
842 |
WirelessHART Process automation device |
843 |
WirelessHART Discrete device |
844 |
WirelessHART Process automation+discrete |
845 |
WirelessHART Gateway |
846 |
WirelessHART Access point |
847 |
WirelessHART Process adapter |
848 |
WirelessHART Discrete adapter |
849 |
WirelessHART enabled handheld/portable maintenance tool |
850 |
Device Profile |
851 |
US |
852 |
United States of America |
853 |
JP |
854 |
Japan |
855 |
DE |
856 |
Germany |
857 |
FR |
858 |
France |
859 |
ES |
860 |
Spain |
861 |
RU |
862 |
Russian Federation |
863 |
CN |
864 |
People's Republic of China |
865 |
Secondary Master |
866 |
Primary Master |
867 |
Device Diagnostic Status 0 |
868 |
Device Diagnostic Status 1 |
869 |
I/O and Subdevice Status |
870 |
WirelessHART Status |
871 |
Time Set Option |
872 |
Real Time Clock Flags |
873 |
Country |
874 |
Country of intended device installation |
875 |
SI Unit Control |
876 |
Indicates if device should only display SI unit codes |
877 |
8.3f |
878 |
.3f |
879 |
Time Stamp |
880 |
Relative time value on network |
881 |
8u |
882 |
Dynamic Variables Returned for Device Variables |
883 |
Command Response Truncated |
884 |
Configuration Change Counter Mismatch |
885 |
Status bytes mismatch |
886 |
Device Variable |
887 |
Reset Modification Flag now? |
888 |
Cancel;Yes |
889 |
Modification Flag is reset! |
890 |
Reset device now? |
891 |
Resetting... %{l_iProcessPercent}% |
892 |
Software reset done. |
893 |
Device is reset!
In order to update all parameter values, disconnect and reconnect the communication to the device. |
894 |
Resetting... |
895 |
Error Log is reset! |
896 |
Reset Modification Log? |
897 |
Modification Log is reset! |
898 |
Safety is already deactivated! |
899 |
Enter Password: |
900 |
Resetting device... |
901 |
Safety Mode is deactivated!
Error Handling set to NORMAL!
Device is not write protected! |
902 |
No Plateau Data available |
903 |
Refreshing ... (Entry No.: %{i_plateau_id}) |
904 |
Refresh Error Log? |
905 |
Refreshing ... %{l_iProcessPercent}% |
906 |
Error Log refresh done. |
907 |
Refresh Modification Log? |
908 |
Modification Log refresh done. |
909 |
Set Key Part1 |
910 |
Set Key Part2 |
911 |
Set Key Part3 |
912 |
Set Key Part4 |
913 |
Set Key Part5 |
914 |
Send Key:
%{0}-%{1}-%{2}-%{3}-%{4} |
915 |
Performing software reset. This will take 15 seconds. |
916 |
Safety is already activated! |
917 |
Safety Mode is activated!
Device is write protected! |
918 |
Device is in write protect mode!
Please disable write protect first! |
919 |
Preparing ReadIn... |
920 |
Select ReadIn Mode: |
921 |
Another ReadIn already running! |
922 |
Error - Invalid Selection |
923 |
ReadIn in process
Rate = %{0} cps
Remaining Time = %{1} s |
924 |
Attention: This will overwrite your measurement settings! |
925 |
Performing calibration ... 0% |
926 |
Performing calibration ... 15% |
927 |
Performing calibration ... 30% |
928 |
Performing calibration ... 45% |
929 |
Performing calibration ... 60% |
930 |
Performing calibration ... 75% |
931 |
Performing calibration ... 90% |
932 |
Performing calibration ... 100% |
933 |
Unknown device status |
934 |
Performing restore ... 0% |
935 |
Performing restore ... 20% |
936 |
Performing restore ... 35% |
937 |
Performing restore ... 50% |
938 |
Performing restore ... 65% |
939 |
Performing restore ... 80% |
940 |
Performing restore ... 100% |
941 |
Restore Status:
%{0}
|
942 |
This will overwrite your actual calibration settings! |
943 |
Performing recall ... 0% |
944 |
Performing recall ... 15% |
945 |
Performing recall ... 30% |
946 |
Performing recall ... 45% |
947 |
Performing recall ... 60% |
948 |
Performing recall ... 75% |
949 |
Performing recall ... 90% |
950 |
Performing recall ... 100% |
951 |
Copy of measuring data transferred to calibration parameter! |
952 |
Enable current loop monitoring now? |
953 |
Yes;No |