Loading…
名称 ID 标签 描述 类型 访问 分类
event1_oper_time_minute 4068 Minutes Shows the time from the overrun logging until now. 无符号 34
event2_oper_time_day 4069 Days Shows the time from the underrun logging until now. 无符号 34
event2_oper_time_hour 406A Hours Shows the time from the underrun logging until now. 无符号 34
event2_oper_time_minute 406B Minutes Shows the time from the underrun logging until now. 无符号 34
float_bloc_byte_var_index 4072 读/写 索引 4096
float_bloc_on_off 406E Float blocking Start/Stop the Float Blocking indication. The function allows the device to recognize a blocked float. 读/写 枚举 2048
control_time_value 4070 Set supervis. time for Select the supervision time. The time can be increased if a quiet flow application is used. 读/写 枚举 2048
low_cut_value 406F Set lower limit Select the supervision range. It can be reduced if a supervision is not possible in the lower measuring range. 读/写 枚举 2048
start_autozero 4071 Start autozero Press F4 to start Autozero recognition. It will take 90 sec to complete. 枚举 2048
autozero_value 4073 Autozero value The Autozero value should be smaller than 0.200. If not repeat function and/or refer to the Instruction Manual. 无符号 2050
名称 ID 标签 描述 类型 访问 分类