Loading…
Name ID Label Description Type Access Class
ramc_long_tag 4035 Long Tag Long Tag- Text that is associated with the Field Device installation. This text can be used by the user in any way. A recommended use is as a unique label to a plant that correlates to a Field Device label: a plant drawing, or on a Control System. R/W ASCII 2048
ramc_ms_code 4037 MS code Model specification code of the measuring tube and indicator R ASCII 2048
templog_byte_var_index 4056 Ninguno R/W Index 4096
templog_on_off 4054 Temp max log Start/stop the temperature logging function. The max temperature value will be logged. R/W Enum 2048
templog_clear 4055 Clear values Copies the logged values to 'Old durat/value', clears the values and resets the temperature to the actual value. W Enum 2048
temp_oper_time_day 4057 Days Shows the time from the temperature logging until now. R Unsigned 34
temp_oper_time_hour 4058 Hours Shows the time from the temperature logging until now. R Unsigned 34
temp_oper_time_minute 4059 Minutes Shows the time from the temperature logging until now. R Unsigned 34
shadow_temp_oper_time_day 405A Days Shows the previous time from the temperature logging until clearing or power off. R Unsigned 2048
shadow_temp_oper_time_hour 405B Hours Shows the previous time from the temperature logging until clearing or power off. R Unsigned 2048
Name ID Label Description Type Access Class