Name ID Label Description Type Access Class
sv_value 4003 SV The conductivity value that changes magnitude with respect to time with no user intervention. R Float 34
temp_value 4013 Temp Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R/W Float 34
temp_ref 4017 Ref.Temp. Reference temperature used for temp. compensation R/W Float 34
ma_output 400A AO1 Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
ma_perc 400B PV % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 66
tc1_perc 4019 TC1 perc Temperature compensation percentage 1 R/W Float 2050
tc2_perc 401A TC2 perc Temperature compensation percentage 2 R/W Float 2050
perc_conc 401B Conc. % Concentration percentage R Float 2050
if_value 401D CC Cell constant R/W Float 2050
sensor_offset 401E Sens.off. Sensor offset (value measured during AIRCAL) R/W Float 2050
Name ID Label Description Type Access Class