Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
bi_dir_mode | 4019 | Bi-dir mode | Bi-directional flow measurement setup- Enables/disables bi-directional flow measurement mode. In this mode, when the pressure equivalent to 0% is applied, the device outputs 50%. Applicable only for differential pressure transmitters. | R/W | Enum | 2048 |
quick_resp | 401A | Quick resp | Quick response- Enables/disables to set the damping value less than 0.5 sec. | R/W | Enum | 2048 |
tz_cmp_mode | 405C | T.Z. Cmp mode | Temperature Zero compensation mode- Enables/disables temperature zero compensation. | R/W | Enum | 2048 |
temp_zero | 405D | Temp Zero | Temperature Zero shift compensation- Defines the value to be used for temperature zero shift compensation. | R/W | Float | 2048 |
sc_enable | 405E | S.C. | Signal characterizer- Enables/disables signal characterizer function. | R/W | Enum | 2048 |
num_of_points | 405F | Num of points | Number of break points- Sets the number of points for signal characterizer. | R/W | Unsigned | 2048 |
x_start | 4061 | X Start | Displays the start point of X. | R | Float | 2048 |
y_start | 4062 | Y Start | Displays the start point of Y. | R | Float | 2048 |
x_end | 4063 | X End | Displays the end point of X. | R | Float | 2048 |
y_end | 4064 | Y End | Displays the end point of Y. | R | Float | 2048 |
Name | ID | Label | Description | Type | Access | Class |