Loading…
Name ID Label Description Type Access Class
device_specific_status_17_latched_value 1287 Status group 10 W BitEnum 2049
device_specific_status_18_latched_value 1288 Latched Device Specific Status 18 W BitEnum 2049
device_specific_status_19_latched_value 1289 Latched Device Specific Status 19 W BitEnum 2049
device_specific_status_20_latched_value 128A Latched Device Specific Status 20 W BitEnum 2049
device_specific_status_21_latched_value 128B Latched Device Specific Status 21 W BitEnum 2049
device_specific_status_22_latched_value 128C Latched Device Specific Status 22 W BitEnum 2049
device_specific_status_23_latched_value 128D Latched Device Specific Status 23 W BitEnum 2049
device_specific_status_24_latched_value 128E Latched Device Specific Status 24 W BitEnum 2049
pressure_units 138D Unit Unidad de variable de dispositivo de campo: Unidad de medición que se mostrará con la representación del valor digital, span mínimo, valor de rango superior, valor de rango inferior, límite inferior del sensor y límite superior del sensor. Presión: Presión del proceso medida con respecto a una presión de referencia R/W Enum 8
pressure_value 138E Pres Valor de variable de dispositivo de campo: Valor digital que cambia de magnitud con respecto al tiempo sin intervención del usuario. Presión: Presión del proceso medida con respecto a una presión de referencia. R Float 10
pressure_percent_range 138F Pres % Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 34
pressure_damping_value 1398 Pres Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. Pressure- Pressure of the process measured with respect to a reference pressure. R/W Float 32
pressure_lower_range_value 1397 LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. Pressure- Pressure of the process measured with respect to a reference pressure. R/W Float 32
pressure_upper_range_value 1396 URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. Pressure- Pressure of the process measured with respect to a reference pressure. R/W Float 32
pressure_lower_range_limit 1394 LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 8
pressure_upper_range_limit 1393 USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 8
pressure_minimum_span 1390 Min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. Pressure- Pressure of the process measured with respect to a reference pressure. R Float 8
pressure_output_transfer_function 1391 Xfer fnctn Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. Pressure- Pressure of the process measured with respect to a reference pressure. R/W Enum 32
module_fill_fluid 139E Fill fluid Sensor Fill Fluid- Fluid material in the module R/W Enum 2048
flange_type 139A Process Conn type Flange Type- Hardware, adjacent to the sensor, that physically connects the process to the sensor R/W Enum 2048
number_of_remote_seals 139F Num of RS Number of Remote Seals- Physical quantity of the seals for the remote seal that is installed R/W Enum 2048
remote_seal_type 139D RS type Remote Seal Type- Type of device that is capable of detecting and passing on the process pressure on to the module. This is only unambiguous if 1 remote seal is installed. R/W Enum 2048
analog_output_value 1779 AO Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes. Analog Output- Measurable physical output which changes with respect to time, without user intervention. R/W Float 66
analog_output_alarm_select 177A AO alm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. Analog Output- Measurable physical output which changes with respect to time, without user intervention. R Enum 1026
module_type_code 177B Module type Measurement Type- Type of measurement application this Field Device is intended to be used with R/W Enum 2048
sensor_a_lower_sensor_trim_point 1774 P LTP Lower Sensor Trim Point- Applied process pressure at which the last low sensor trim was performed. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R Float 8
sensor_a_upper_sensor_trim_point 1776 P UTP Upper Sensor Trim Point- Applied process pressure at which the last upper sensor trim was performed. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R Float 8
flange_material 177C Process Conn matl Flange Material- Construction material of flange type R/W Enum 2048
remote_seal_isolator_material 177E RS isoltr matl Remote Seal Isolator Material- Construction material of a part of the remote seal type that comes in contact with the process material. This is only unambiguous if 1 remote seal is installed. R/W Enum 2048
module_isolator_material 177F Isoltr matl Sensor Isolator Material- Construction material of a part of the module type that comes in contact with the process material R/W Enum 2048
drain_vent_material 1780 Drain vent matl Drain Vent Material- Construction material of a removable plug that exists on the flange type that can be opened temporarily to remove unwanted process from the sensor area R/W Enum 2048
pressure_sensor_serial_number 1770 Pres snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R Unsigned 2048
Name ID Label Description Type Access Class