Name ID Label Description Type Access Class
flow_units 1F7B PV unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 32
flow_value 1F7C PV Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 34
flow_analog_value 1F7D PV AO Valore uscita analogica - Valore che traccia la rappresentazione del valore digitale, sotto le modalità operative normali R/W Float 34
flow_percent_range 1F7E PV % rnge Percentuale di campo - Variabile che traccia la rappresentazione del valore digitale rispetto al campo definito dal valore di campo inferiore e dal valore di campo superiore, per le modalità operative normali. Le unità di questa variabile sono sempre in percentuale. R Float 34
dummy_float_value1 4003 Dummy float value R Float 32
dummy_float_value2 4004 Dummy float value R Float 32
dummy_float_value3 4005 Dummy float value R Float 32
dummy_float_value4 4009 Dummy float value R/W Float 32
dummy_3integer_value 4006 Dummy Integer3 Value R Unsigned 2048
flow_alarm_select 1F83 4-20mA Burn Out Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R Enum 64
dummy_transfer_func 4007 Dummy transfer function R Enum 2048
flow_span_disp 4008 PV Span Sets flow rate span (with units from "Base Flow Unit" and "Base Time Unit"). R/W Float 32
flow_damping_value 1F87 PV Damping Valore di smorzamento - Costante di tempo di smorzamento applicata sia alla rappresentazione del valore digitale che al valore analogico della rispettiva uscita analogica. Le unità di questa variabile sono sempre in secondi. R/W Float 32
flow_total_units 400B Total Unit R Enum 32
flow_total_value 400C Totl Displays the totalized value for flow rate in the forward direction. R Float 34
rev_total_units 400E Reverse Totl Units R Enum 32
rev_total_value 400F Reverse Totl Displays the totalized value for flow rate in the reverse direction. R Float 34
dif_total_units 4010 Dif Totl Units R Enum 32
dif_total_value 4011 Dif Totl Displays the totalized value for flow rate between forward totalization and reverse totalization. R Float 34
select_flow_tube 4012 Select Flow Tube Selects the flowtube's model name. R/W Enum 2048
nominal_size_unit 4013 Nominal Size Unit Selects the nominal size units for the flowtube. R/W Enum 2048
nominal_size 4014 Nominal Size Sets flowtube nominal size in selected unit at "Nominal Size Unit". R/W Float 2048
measure_mode 4015 Measure Mode Selects measurement mode for dual frequency excitation. R/W Enum 2048
low_mf 4016 Low MF Sets low-frequency meter factor for standard dual frequency excitation. R/W Float 2048
high_mf 4017 High MF Sets high-frequency meter factor for standard dual frequency excitation. R/W Float 2048
low_mf_edf 4018 Low MF(EDF) Sets low-frequency meter factor for enhanced dual frequency excitation. R/W Float 2048
high_mf_edf 4019 High MF(EDF) Sets high-frequency meter factor for enhanced dual frequency excitation. R/W Float 2048
span_unit 401A Span Unit R Enum 2048
pv_span 401B PV Span Sets flow rate span (with units from "Base Flow Unit" and "Base Time Unit"). R/W Float 32
base_flow_unit 401C Base Flow Unit Selects flow units for the flow rate span. R/W Enum 2048
base_time_unit 401D Base Time Unit Selects time units for the flow rate span. R/W Enum 2048
flow_decimal_point 401E Flow Decimal Pnt Selects decimal point position for the display unit's instantaneous flow rate. R/W Enum 2048
density_unit 401F Density Unit Sets units for density when mass flow rate is selected. R/W Enum 2048
mass_flow_density 4020 Mass Flow Density Sets density when mass flow rate is selected (with units from "Density Unit"). R/W Float 2048
user_span_select 4021 User Span Select Selects whether or not special units are used as flow rate units. R/W Enum 2048
user_flow_unit 4022 Flow User Unit Sets the special flow rate units. R/W ASCII 2048
user_flow_span 4023 Flow User Span Sets span when using special flow rate units. R/W Float 2048
analog_low_cut 4024 4-20mA Low Cut Sets the range in vicinity of 0 % within which the current output will be 4 mA. R/W Float 2048
analog_low_lmt 4025 4-20mA Low Lmt Sets the low limit for current output. R/W Float 2048
analog_high_lmt 4026 4-20mA High Lmt Sets the high limit for current output. R/W Float 2048
velocity_check 4027 Velocity Check Display of the span setting using flow velocity (m/s). R Float 2050
magflow_zero 4028 Magflow Zero Displays the result of the automatic zero adjustment, or sets the zero point. R/W Float 2048
autozero_time 4029 Autozero Time R/W Integer 2048
auto_zero_exe 402A Auto Zero Exe Selects whether or not automatic zero adjustment is carried out. R Enum 2048
disp_select1 402B Display Select1 Selects content of the first line for Display Mode. R/W Enum 2048
disp_select2 402C Display Select2 Selects content of the second line for Display Mode. R/W Enum 2048
disp_select3 402D Display Select3 Selects content of the third line for Display Mode. R/W Enum 2048
disp_cycle 402E Display Cycle Selects the display cycle. R/W Enum 2048
lcd_contrast 402F LCD Contrast R/W Integer 2048
language 4030 Language Selects the language used for the display unit. R/W Enum 2048
Name ID Label Description Type Access Class