Name ID Label Description Type Access Class
sw_position 40FE SW position Setting the switch position enables the display orientation to be kept fixed, regardless of in which direction relative to the piping the converter is installed. R/W Enum 2048
input_range_type 4015 Type SINGLE:Single range. 4F-0R:Unidirectional flow, automatic selection of multiple ranges. 2F-2R:Bidirectional flows, automatic selection of multiple ranges. EXT 2F-0R:Unidirectional flow, multiple ranges selected by external signal. EXT 2F-2R:Bidirectional flows, multiple ranges selected by external signal. R/W Enum 32
input_range_units 400F Unit Range Unit- Engineering unit to be displayed with the Upper Range Value and Lower Range Value R/W Enum 32
input_upper_value 4010 Range1 Range1 span value. Range can be set within 0.1m/s to 10m/s in terms of flow velocity. R/W Float 32
input_upper_value2 4011 Range2 Range2 span value. R/W Float 32
input_upper_value3 4012 Range3 Range3 span value. R/W Float 32
input_upper_value4 4013 Range4 Range4 span value. R/W Float 32
input_multi_range_hysteresys 4016 Range hys Range hysteresis is the dead band used when multiple ranges are switched. You can set the hysteresis within the range of 0 to 25% in increments of 0.1%. R/W Unsigned 4
fixed_pulse 40FF Fixed pulse R/W Unsigned 32
fixed_current 4100 Fixed current R/W Float 32
Name ID Label Description Type Access Class