Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
input_pulse_width | 4019 | Pulse width | R/W | Unsigned | 32 | |
input_pulse_mode | 40F2 | Pulse mode | AUTO:After the count rate is set, the pulse width is automatically set to 40% of the period of pulse frequency at 100% output. MANUAL:Even after the count rate is set, the pulse width is not changed. | R/W | Enum | 32 |
input_d_size | 401C | Meter size | R/W | Unsigned | 32 | |
input_d_size_unit | 40AA | Meter size unit | R/W | Enum | 32 | |
input_signal_status_code | 401F | Alarm set1 | R/W | BitEnum | 32 | |
input_signal_status_code2 | 40F3 | Alarm set2 | R/W | BitEnum | 32 | |
input_alarm_upper_value | 401D | High alarm | R/W | Integer | 32 | |
input_alarm_lower_value | 401E | Low alarm | R/W | Integer | 32 | |
input_empty_level | 40AB | Empty alarm level | R/W | Enum | 32 | |
input_alarm_upper_value2 | 40F4 | HH alarm | R/W | Integer | 32 | |
input_alarm_lower_value2 | 40F5 | LL alarm | R/W | Integer | 32 | |
input_signal_percent_rate_limit | 4020 | Rate-of-change limit | R/W | Unsigned | 32 | |
input_signal_time_rate_limit | 4021 | Control limit time | R/W | Unsigned | 32 | |
sensor_damping_value | 4009 | Damping | Damping value is used to moderate output fluctuations. The larger damping value, the more the output is averaged. But the response to an input change will be slower.Damping value is set for 0.0sec, 0.5sec and 1 to 60sec (in increments of 1sec). | R/W | Float | 8 |
detector_serial_number | 4023 | Detector | Detector's serial number | R | PackedAscii | 2048 |
converter_serial_number | 4024 | Converter | Converter's serial number | R | PackedAscii | 2048 |
detector_spec_code | 4090 | Detector | Detector's product specification code | R | PackedAscii | 2048 |
converter_spec_code | 4091 | Converter | Converter's product specification code | R | PackedAscii | 2048 |
input_ex_curr | 408D | Exc current | Be sure to match the exciting current value with the value specified for the combined detector. Specifying any other value may cause an error. The setting range of excitation current value is from 0.0000A to 0.2500A. | R/W | Float | 32 |
input_flange_types | 4092 | Flange type | R/W | Enum | 2048 | |
LF600_liner_material | 40F6 | Lining material | R/W | Enum | 2048 | |
LF600_electrode_material | 40F7 | Electrode material | R/W | Enum | 2048 | |
connection_flgn_matl | 402C | Earthring material | R/W | Enum | 2048 | |
seal_material | 4029 | Seal matl | R/W | Enum | 2048 | |
input_pass_word | 40B4 | Password | R/W | Unsigned | 32 | |
Name | ID | Label | Description | Type | Access | Class |