Loading…
Name ID Label Description Type Access Class
hardware_revision 009F Hardware rev Hardware Revision- Revision that corresponds to the electronics hardware of the Field Device. R Unsigned 2048
physical_signaling_code 00AB Physicl signl code Physical Signaling Code- The Physical Signaling Code describes the type of Physical Layer that has been implemented in the hardware that is responsible for the HART Communication port R Enum 2048
device_flags 00A0 Dev flags Field Device Flags- Describes special features of the Field Device that affect the Data Link Layer R BitEnum 128
device_id 00A1 Dev id Field Device Identification- Uniquely identifies the Field Device when combined with the Manufacturer Identification and Device Type. Therefore, this variable cannot be modified by the Host user. R Unsigned 128
polling_address 00A2 Poll addr Polling Address- Address used by the Host to identify a Field Device, and changeable by the User to control: the Multidrop Mode for Analog Output 1), and the Analog Output Fixed Mode of Analog Output 1) W Unsigned 128
tag 00A3 Tag Tag- Text that is associated with the Field Device installation. This text can be used by the user in any way. A recommended use is as a unique label to a plant that correlates to a Field Device label: a plant drawing, or on a Control System. This variable is also used as a type of data link layer address handle. R/W PackedAscii 2048
message 00A4 Message Message- Text that is associated with the Field Device. This text can be used by the user in any way. There is no recommended use. R/W PackedAscii 2048
descriptor 00A5 Descriptor Descriptor- Text that is associated with the Field Device. This text can be used by the user in any way. There is no specific recommended use. R/W PackedAscii 2048
date 00A6 Date Date- Gregorian calendar date that is stored in the Field Device. This date can be used by the user in any way. There is no specific recommended use. R/W HartDate 2048
write_protect 00A7 Write protect Write Protect Mode- Indicates whether variables can be written to the Field Device, or whether commands that cause actions, to be performed in the Field Device, can or cannot occur. Depending on the Field Device, this is a mode that is controllable by the user, either via the Field Device hardware or the HART Protocol. R Enum 2048
private_label_distributor 00A8 Distributor Private Label Distributor- References the company that is responsible for the distribution of this Field Device to customers R Enum 2048
final_assembly_number 00A9 Final asmbly num Final Assembly Number- Number that is used for identification purposes, and is associated with the overall Field Device. R/W Unsigned 2048
xmtr_specific_status_0 0FB5 Status group 1 Field Device Specific Status Byte 1- Data that has Field Device status information relating to: failures / warnings / status of process's. R BitEnum 2051
xmtr_specific_status_1 0FB6 Status group 2 Field Device Specific Status Byte 2- Data that has Field Device status information relating to: failures / warnings / status of process's. R BitEnum 2051
xmtr_specific_status_2 0FB7 Status group 3 Field Device Specific Status Byte 3- Data that has Field Device status information relating to: failures / warnings / status of process's. R BitEnum 2051
xmtr_specific_status_3 0FB8 Status group 4 Field Device Specific Status Byte 4- Data that has Field Device status information relating to: failures / warnings / status of process's. R BitEnum 2051
flange_type 139A Flange type Flange Type- Hardware, adjacent to the sensor, that physically connects the process to the sensor R/W Enum 2048
o_ring_material 139B O ring matl O Ring Material- Construction material of a seal that exists between the module and flange R/W Enum 2048
meter_installation 139C Meter type Meter Installation- Defines whether there is a meter installed or not and possibly what information the meter will display R/W Enum 256
remote_seal_type 139D RS type Remote Seal Type- Type of device that is capable of detecting and passing on the process pressure on to the module. This is only unambiguous if 1 remote seal is installed. R/W Enum 2048
module_fill_fluid 139E Fill fluid Sensor Fill Fluid- Fluid material in the module R Enum 2048
number_of_remote_seals 139F Num remote seal Number of Remote Seals- Physical quantity of the seals for the remote seal that is installed R/W Enum 2048
sensor_a_lower_sensor_trim_point 1774 Pres snsr LST pt Lower Sensor Trim Point- Applied process pressure at which the last low sensor trim was performed. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R/W Float 8
sensor_a_upper_sensor_trim_point 1776 Pres snsr UST pt Upper Sensor Trim Point- Applied process pressure at which the last upper sensor trim was performed. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R/W Float 8
sensor_a_sensor_trim_unit 1772 Pres snsr ST unit Sensor Trim Unit- Unit associated with the Sensor Lower Trim and Sensor Upper Trim. Pressure Sensor- Pressure hardware that measures pressure with respect to a reference pressure. R Enum 8
analog_output_value 1779 AO1 Out Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes. Analog Output- Measurable physical output which changes with respect to time, without user intervention. R/W Float 66
analog_output_alarm_select 177A AO1 Alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. Analog Output- Measurable physical output which changes with respect to time, without user intervention. R Enum 64
module_type_code 177B Meas typ Measurement Type- Type of measurement application this Field Device is intended to be used with R Enum 2048
flange_material 177C Flnge matl Flange Material- Construction material of flange type R/W Enum 2048
meter_information_configuration 177D Meter typ config Meter Information Configuration- Defines what information the meter will display R/W Enum 256
remote_seal_isolator_material 177E RS isoltr matl Remote Seal Isolator Material- Construction material of a part of the remote seal type that comes in contact with the process material. This is only unambiguous if 1 remote seal is installed. R/W Enum 2048
module_isolator_material 177F Isoltr matl Sensor Isolator Material- Construction material of a part of the module type that comes in contact with the process material R Enum 2048
drain_vent_material 1780 Drain vent matl Drain Vent Material- Construction material of a removable plug that exists on the flange type that can be opened temporarily to remove unwanted process from the sensor area R/W Enum 2048
Name ID Label Description Type Access Class