Loading…
Name ID Label Description Type Access Class
DiffPressMaximumUTP 4014 DP Max upper trim Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0
DiffPressMinimumTrimDiff 4015 DP Min trim diff Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. R Float 0
DiffPressSimulated 4016 DP Simulate W Enum 2052
DiffPress_upperRange_value 4017 PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
DiffPress_lowerRange_value 4018 PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
DiffPressCalStatus 4019 Cal Status DP Input Calibration Status R/W Enum 2048
DiffPressCalCount 401A Cal. Index Counter DP Input Calibration Index Count R Unsigned 2048
DiffPressCalEUZO_value 401B EU ZO Val Cal. Pt #1 EU Zero Value R/W Float 2048
DiffPressCalRawZO_value 401C Raw ZO Val Cal. Pt #1 Raw Zero Value R/W Float 2048
DiffPressCalEUMid_value 401D EU Mid Val Cal. Pt #2 EU Mid Value R/W Float 2048
Name ID Label Description Type Access Class