Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
RTDTempMaximumUTP | 409A | RTDTemp Max upper trim | Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 0 |
RTDTempMinimumTrimDiff | 409B | RTDTemp Min trim diff | Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. | R | Float | 0 |
RTDTempSimulated | 409C | RTDTemp Simulate | W | Enum | 2052 | |
RTDTemp_upperRange_value | 409D | PV URV | Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
RTDTemp_lowerRange_value | 409E | PV LRV | Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 32 |
RTDTempCalStatus | 409F | Cal Status | RTD Temp Input Calibration Status | R/W | Enum | 2048 |
RTDTempCalCount | 40A0 | Cal. Index Counter | RTD Temperature Input Calibration Index Count | R | Unsigned | 2048 |
RTDTempCalEUZO_value | 40A1 | EU ZO Val | RTD Temperature Cal. Pt #1 EU Zero Value | R/W | Float | 2048 |
RTDTempCalRawZO_value | 40A2 | Raw ZO Val | RTD Temperature Cal. Pt #1 Raw Zero Value | R/W | Float | 2048 |
RTDTempCalEUMid_value | 40A3 | EU Mid Val | RTD Temperature Cal. Pt #1 EU Mid Value | R/W | Float | 2048 |