Name ID Label Description Type Access Class
OffsetManSensorCalibration 4157 Man. Corr. Offset (Fct. D3.3.1) R/W Float 0
OffsetMidSensorCalibration 4158 Corr. Offset Mid (Fct. D3.3.5) R Float 0
OffsetReference 4159 Reference Offset Reference Offset allows shifting the distance output, e.g. to compensate different installation heights across a facility (Fct. C1.10) R/W Float 0
OffsetSensorCalibration 415A Corr. Offset (Fct. D3.3.3) R Float 0
OffsetTankBottom 415B Tank Bottom Offset Tank Bottom Offset allows shifting the level output value, e.g. to compensate different installation heights across a facility (Fct. C1.11) R/W Float 0
OperatingTime 415C Operating Time [s] Total operating time of the device in seconds. (Fct. B2.1, D4.1.1) R Unsigned 0
OptionEmptySpectrum 415D Empty Spectrum Enable Empty Spectrum mode can be used to improve results by removing unwanted reflections from installations (ladders, struts, nozzles etc) (Fct. C2.8) R/W Enum 0
OptionLinesDisplayPage1 415E Function Select the measurement page type (Fct. C6.4.1) R/W Enum 0
OptionLinesDisplayPage2 415F Function Select the measurement page type (Fct. C6.5.1) R/W Enum 0
OptionMeasurementDisplayLine1Page1 4160 1st Value Variable Select the variable for the first measurement line (Fct. C6.4.2) R/W Enum 0