Loading…
Name ID Label Description Type Access Class
flow_damping_value 1F87 PV Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
pulse_output_value 1F73 Puls outp Pulse Output Value- scaled frequency output from 0-1000 Hz, linear with input. R Float 514
pulse_output_test 1F66 Puls outp test Pulse Output Test Value- Scaled frequency output from 0-1000 Hz, linear with input. R/W Integer 4096
gross_total_value 1F89 Measure gross totl Gross Total Value- Non-resettable totalizer value. Counts frequency output pulses while total control is turned on. This value will be reset to 0 when the pulse scaling, line size, or units are changed. R Float 18
net_total_value 1F8A Measure net totl Net Total Value- Resettable totalizer value. Counts frequency output pulses while total control is turned on. R Float 18
reverse_total_value 1F8B Measure reverse totl Reverse Total Value- Non-resettable totalizer value. Counts frequency output pulses while total control is turned on and flow is in the reverse direction. This value will be reset to 0 when the pulse scaling, line size, or units are changed. R Float 18
unique_device_type_code 07ED R/W Enum 128
rosemount_model_code 07F5 R/W Enum 2048
rosemount_analytical_model_code 07F6 R/W Enum 2048
kayray_model_code 07F7 R/W Enum 2048
Name ID Label Description Type Access Class