Loading…
Name ID Label Description Type Access Class
ST_RjType_2 408B Reference junction 2 Selection of reference junction measurement for temperature compensation in thermocouples. R/W Enum 2048
ST_FixedRjValue_2 408C RJ preset value 2 Setting of the value for the external temperature compensation. R/W Float 2048
ST_CalVandusenA_2 408D Call./v. Dusen coeff. A Setting of the Callendar/van Dusen coefficient A for the linearization based on the Callendar-Van Dusen method. R/W Float 2048
ST_CalVandusenB_2 408E Call./v. Dusen coeff. B Setting of the Callendar/van Dusen coefficient B for the linearization based on the Callendar-Van Dusen method. R/W Float 2048
ST_CalVandusenC_2 408F Call./v. Dusen coeff. C Setting of the Callendar/van Dusen coefficient C for the linearization based on the Callendar-Van Dusen method. R/W Float 2048
ST_CalVandusenR0_2 4090 Call./v. Dusen coeff. R0 Setting of the R0 value for the linearization based on the Callendar-Van Dusen method. R/W Float 2048
ST_PolyNickelCopperA_2 4091 Polynomial coeff. A Setting of the Polynomial coefficient A for Sensor linearization of copper/nickel resistance thermometers (RTD). R/W Float 2048
ST_PolyNickelCopperB_2 4092 Polynomial coeff. B Setting of the Polynomial coefficient B for Sensor linearization of copper/nickel resistance thermometers (RTD). R/W Float 2048
ST_PolyNickelCopperR0_2 4093 Polynomial coeff. R0 Setting of the R0 value for Sensor linearization of copper/nickel resistance thermometers (RTD). R/W Float 2048
ST_SensorTrimHi_2 4094 Sensor trimming upper value Setting of the upper point for linear characteristic calibration (this affects offset of and slope). R/W Float 2048
Name ID Label Description Type Access Class