Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
pvValue | 4001 | Gas | Field Device Variable Value | R | Float | 2 |
pvUnits | 4002 | PV unit | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | R/W | Enum | 0 |
pvFamily | 4003 | PV Sensor Family | Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. | R | Enum | 0 |
pvClassification | 4004 | PV Sensor Class | Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. | R | Enum | 0 |
pvPDQ | 4005 | Gas PDQ | Process Data Quality- Indicates the overall process data quality of the device variable value. | R | Enum | 2 |
pvLS | 4006 | Gas LS | Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). | R | Enum | 2 |
pvStatus | 4007 | Gas Fmly stat | Device Family Status- Indicates additional status information on a device variable for a specific device family. | R | BitEnum | 2 |
pvUSL | 4008 | PV USL | Upper Sensor Limit Value- Defines the maximum measurable value for the Upper Range Value | R | Float | 0 |
pvLSL | 4009 | PV LSL | Upper Sensor Limit Value- Defines the maximum measurable value for the Lower Range Value | R | Float | 0 |
pvMinimumSpan | 400A | PV Min span | Minimum Span Value- Defines the difference value between the Upper Range Value and the Lower Range Value. | R | Float | 0 |