Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
dvUpperTrimPoint4 | 40E0 | Pres Upper trim | Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. | R | Float | 0 |
dvTrimSupport4 | 40E1 | Pres Trim support | Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. | R | Enum | 0 |
dvMinimumLTP4 | 40E2 | Pres Min lower trim | Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 0 |
dvMaximumLTP4 | 40E3 | Pres Max lower trim | Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 0 |
dvMinimumUTP4 | 40E4 | Pres Min upper trim | Minimum Upper Trim Point- The lowest applied process value that can be used duirng a high trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 0 |
dvMaximumUTP4 | 40E5 | Pres Max upper trim | Maximum Upper Trim Point- The highest applied process value that can be used duirng a high trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 0 |
dvMinimumTrimDiff4 | 40E6 | Pres Min trim diff | Minimum Trim Differential- The minimum acceptable difference between the upper and lower trim points. | R | Float | 0 |
dvTrimMode4 | 40E8 | Trim Mode | Trimming increases the measurement accuracy of the instrument, any point between 0% and 100% along the scale can be trim. Note one-point trimming applies an offset to the sensor reading, while two-point trimming applies both an offset and a gain. | R/W | Enum | 8 |
dvLowerTrimValue4 | 40E9 | dvLowerTrimValue | Lower Trim Point- Applied process input at which the last lower trim of the device variable was performed. | R | Float | 0 |
dvUpperTrimValue4 | 40EA | dvUpperTrimValue | Upper Trim Point- Applied process input at which the last upper trim of the device variable was performed. | R | Float | 0 |
Name | ID | Label | Description | Type | Access | Class |