Name ID Label Description Type Access Class
primary_measure_upper_sensor_limit 400C USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
primary_measure_lower_sensor_limit 400D LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
trim_mode 403C Trim Mode Input trimming increases the measurement accuracy of the instrument, any point between 0% and 100% along the scale can be trim. Note one-point trimming applies an offset to the sensor reading, while two-point trimming applies both an offset and a gain. R/W Enum 8
primary_measure_minimum_span 400E Min Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
upper_trim_point 4011 UTP Upper trim point value R/W Float 8
lower_trim_point 4010 LTP Lower trim point value R/W Float 8
trim_point_value 4018 TPV Trim Point Value R/W Float 8
TrimSupport 4012 Trim support Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. R Enum 0
MinimumLTP 4013 Min lower trim Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). R Float 0
MaximumLTP 4014 Max lower trim Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). R Float 0