Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
primary_measure_upper_sensor_limit | 400C | USL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 8 |
primary_measure_lower_sensor_limit | 400D | LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. | R | Float | 8 |
trim_mode | 403C | Trim Mode | Input trimming increases the measurement accuracy of the instrument, any point between 0% and 100% along the scale can be trim. Note one-point trimming applies an offset to the sensor reading, while two-point trimming applies both an offset and a gain. | R/W | Enum | 8 |
primary_measure_minimum_span | 400E | Min Span | Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. | R | Float | 8 |
upper_trim_point | 4011 | UTP | Upper trim point value | R/W | Float | 8 |
lower_trim_point | 4010 | LTP | Lower trim point value | R/W | Float | 8 |
trim_point_value | 4018 | TPV | Trim Point Value | R/W | Float | 8 |
TrimSupport | 4012 | Trim support | Trim Points Supported- Indicates the trim points that are supported by the device variable during a trim procedure. | R | Enum | 0 |
MinimumLTP | 4013 | Min lower trim | Minimum Lower Trim Point- The lowest applied process value that can be used duirng a low trim procedure (ie, no value lower than this will be accepted by the instrument). | R | Float | 0 |
MaximumLTP | 4014 | Max lower trim | Maximum Lower Trim Point- The highest applied process value that can be used duirng a low trim procedure (ie, no value higher than this will be accepted by the instrument). | R | Float | 0 |