名称 ID 标签 描述 类型 访问 分类
KDGM_pulse_repetition_rate 4040 Pulse Rate P041 Pulse Rate - the interval between two ultrasonic transmit cycles may be set to, say, avoid crosstalk between two adjacent transmitters. The setting is typically 1.0 second. 读/写 枚举 8
KDGM_echoes_needed 4041 Echoes Needed P042 Echoes Needed - the number of echoes which must be returned from the same target before it is considered to be a valid target. The setting is typically 4 echoes. 读/写 无符号 8
KDGM_target_threshold_1_time 4085 Threshold 1 Time P043 Threshold 1 Time - the threshold may be set to an initial higher level set by P048. Threshold 1 Time sets the duration in milliseconds of that higher threshold. Note that 6 milliseconds is typically equivalent to a distance of 1 metre. 读/写 浮点 8
KDGM_target_number_of_pulses 4042 Target Pulses P044 Number of pulses - the number of cycles in the transmitted ultrasonic burst. The typical value is 10 cycles. 读/写 枚举 8
KDGM_target_frequency_value 402A Target Frequency P045 Target frequency - if set to Automatic the transmitter selects the best frequency for the application. It can be fixed, typically in the range of 48 to 58kHz. Not all integer values are available. 读/写 枚举 8
SM_spike_rejection 4047 Spike Reject P049 Spike Reject - rejects narrow echoes. Increasing P049 increases the width of rejected echoes. The typical value is 0 - i.e. no spike rejection 读/写 无符号 8
SM_auto_cycle 404D Simulation Mode Simulation Mode - the transmitter can simulate the level ramping between minimum and maximum. A complete cycle takes approximately 100 seconds. Auto-cycling may be paused. 读/写 枚举 1024
SM_base_units 402D Base Units Base Units - the transmitter can operate using Base Units of metres, feet or inches. Changing the Base Units will RESET the transmitter and restore it to ex-factory settings in the selected Base Units. 读/写 枚举 2048
MOB_target_threshold_1_size 40BD Threshold 1 Size P048 Threshold 1 Size - the amplitude of the initial high threshold. The duration is set by P043. 读/写 浮点 8
MOB_transmit_power_control 40BE Power Control P040 Transmit Power Control - if enabled then the transmitter automatically adjusts the transmitted pulse energy for optimum accuracy. 读/写 枚举 1024