Name ID Label Description Type Access Class
temperature_value 4013 Xducer Temp D903 Temperature - The value derived from the sensor in the transducer or connected externally (if fitted) R Float 34
temperature_value_unit 4012 Temperature units tvu R Enum 2048
current_output 4006 Current output D906 Current output - The analog value related to the PV R Float 66
percent_range 4007 % Current Out D905 Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
SM_relay_status 4027 Relay Status D908 Relay Status - The status of the transmitter relay contacts. 1 = Closed, 0 = Open R Enum 34
level_damping_value 4008 Damping P020 Smoothing time - The damping in seconds applied to the PV value and the analogue output R/W Float 1024
KDGM_lost_echo_delay 4030 LE Delay P021 Lost echo delay - The delay after which a continuous lost echo is reported, and the action specified by P022 taken R/W Float 64
KDGM_lost_echo_action 4031 LE Action P022 Lost echo action - The value to which the PV and current will be set, if the echo is lost for the time set by P021 R/W Enum 64
KDGM_target_blanking_distance 4032 Blanking P023 Blanking distance - The distance from the sensor up to which all echoes will be ignored R/W Float 8
KDGM_speed_of_sound 4033 Speed of sound P024 Speed of sound - The nominal speed of sound for the ullage space at 0øC R/W Float 8
Name ID Label Description Type Access Class