Loading…
Loading…
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
initial_zero_ref_at_1MHz | 4023 | Initial Zero | Initial Zero Reference @ Frequency A - P123: At first installation the attenuation of the signal by clear supernatant at a frequency of 1MHz (or 3.7MHz) measured in units of dB. | 读/写 | 浮点 | 8 |
initial_zero_ref_at_3MHz | 4024 | Initial Zero | Initial Zero Reference @ Frequency B - P124: At first installation the attenuation of the signal by clear supernatant at a frequency of 3.3MHz measured in units of dB. | 读/写 | 浮点 | 8 |
KDGM_type | 4049 | Model | Model Code - D750: The basic model number of the Control Unit. | 读/写 | 枚举 | 32 |
span_1_calibration_attenuation_at_1MHz | 4027 | Span 1 | Span 1 @ Frequency A - P130: The first of the calibration readings using the sludge at a frequency of 1MHz (or 3.7MHz) measured in units of dB. | 读/写 | 浮点 | 8 |
span_1_calibration_attenuation_at_3MHz | 4028 | Span 1 | Span 1 @ Frequency B - P131: The first of the calibration readings using the sludge at a frequency of 3.3MHz measured in units of dB. | 读/写 | 浮点 | 8 |
span_2_calibration_attenuation_at_1MHz | 4029 | Span 2 | Span 2 @ Frequency A - P132: The second of the calibration readings using the sludge at a frequency of 1MHz (or 3.7MHz) measured in units of dB. | 读/写 | 浮点 | 8 |
span_2_calibration_attenuation_at_3MHz | 402A | Span 2 | Span 2 @ Frequency B - P133: The second of the calibration readings using the sludge at a frequency of 3.3MHz measured in units of dB. | 读/写 | 浮点 | 8 |
span_3_calibration_attenuation_at_1MHz | 402B | Span 3 | Span 3 @ Frequency A - P134: The third of the calibration readings using the sludge at a frequency of 1MHz (or 3.7MHz) measured in units of dB. | 读/写 | 浮点 | 8 |
span_3_calibration_attenuation_at_3MHz | 402C | Span 3 | Span 3 @ Frequency B - P135: The third of the calibration readings using the sludge at a frequency of 3.3MHz measured in units of dB. | 读/写 | 浮点 | 8 |
totaliser_value | 4066 | Totaliser Value | TBD | 读 | 浮点 | 3 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |