Loading…
Name ID Label Description Type Access Class
pv_upper_range_value 400E Upper Range Value Upper Range Value - P401: Upper Range Value - the PV at which the current output is 20 mA and the % output is 100%. R/W Float 32
pv_lower_range_value 400F Lower Range Value Lower Range Value - P400: Lower Range Value - the PV at which the current output is 4 mA (or 0 mA) and the % output is 0%. R/W Float 32
alarm_code 4010 Analog Output Alarm Type R Enum 2048
trans_func 4011 Função Xfer R Enum 2048
sludge_density_units 4012 Density Units R/W Enum 2048
sludge_density 4013 Sludge Density Sludge Density - D801: the density of the medium as determined by the system. R Float 34
attenuation_unit 4014 Attenuation Unit R Enum 2048
attenuation_value 4015 Attenuation Attenuation - D802: the attenuation of the signal by the medium relative to the zero reference calibration point. R Float 34
total_attenuation_unit 4016 Total Attenuation Unit R Enum 2048
total_attenuation_value 4017 Total Attenuation Total Attenuation - D803: the total attenuation of the signal by the medium and associated circuitry. R Float 34
Name ID Label Description Type Access Class