Loading…
Name ID Label Description Type Access Class
span_at_1mhz 4085 Span Span @ Frequency A - D854: The average of the span calibrations at 1MHz (or 3.7MHz) sensor frequency - measured in units of dB. R Float 3
span_at_3mhz 4086 Span Span @ Frequency B - D855: The average of the span calibrations at 3.3MHz sensor frequency - measured in units of dB. R Float 3
slope_at_1mhz 4087 Slope Slope A - D858: The calibrated value of attenuation in the medium at 1MHz (or 3.7MHz) in units of dB/%. R Float 3
slope_at_3mhz 4088 Slope Slope B - D859: The calibrated value of attenuation in the medium at 3.3MHz in units of dB/%. R Float 3
digital_input_status 408A Digital Input Status Digital input status - D835: The status of the control unit Digital Inputs. OFF indicates that the Digital input is inactive and ON indicates that the Digital input is active. A digital input may be used to initiate a desludging operation. R/W BitEnum 3
current_input 408B Current Input TBD R Float 3
start_on 4044 Start On Start On - P250: The event or condition that will initiate a desludging sequence. This may typically be a Time of day, contact closure or measured density value. This parameter and associated timing parameters are only editable locally on the Control Unit. R/W Enum 64
stop_on 4045 Stop On Stop On - P251: The event or condition that will terminate a desludging sequence. This may typically be a Time Interval, a contact closure or measured density value. This parameter and associated timing parameters are only editable locally on the Control Unit. R/W Enum 64
stop_if 4046 Stop If Stop If - P252: The event or condition that will terminate a desludging sequence early. This may typically be a contact closure or measured density value. This parameter and associated timing parameters are only editable locally on the Control Unit. R/W Enum 64
start_time_1 408E Start Time 1 Start Time 1 - P253: If starting a desludge sequence based on Time of Day, this is the Time at which the sequence will start. If set to 00:00 then that start time is not used. R/W Integer 64
Name ID Label Description Type Access Class