Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
fo_fault_level | 43A1 | Fault Level | This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. | R/W | Float | 64 |
density_for_cal | 406B | R/W | Float | 4096 | ||
point_code | 406C | R/W | Unsigned | 4096 | ||
densityfd_for_cal | 406E | R/W | Float | 4096 | ||
temp_for_cal | 4373 | Ref temperature | User-entered temperature for the purpose of performing an on-line temperature calibration. | R/W | Float | 4096 |
D1 | 4089 | D1 | The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D2 | 408B | D2 | The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D3 | 4073 | D3 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
D4 | 4075 | D4 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
K1 | 408A | K1 | The tube period, in microseconds, corresponding to a density equal to D1. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |