Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
vol_flow_digital_units | 4391 | Volume Flow Unit | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. | R/W | Enum | 32 |
vol_flow_digital_value | 41D1 | Flow Rate | R | Float | 10 | |
vol_flow_digital_cutoff | 4552 | Volume Flow Cutoff | The volume flow rate below which the frequency/pulse output indicates zero flow, and below which totalizers stop counting. Cutoffs can also be configured for the mA output(s) (the AO cutoffs). An AO cutoff is active only if the process variable assigned to the output is mass flow or volume flow. | R/W | Float | 8 |
vol_flow_analog_value | 464E | Volume Flow mA Value | R/W | Float | 66 | |
vol_flow_analog_upper_limit | 469D | アナログ Upper リミット | アナログ アウトプット アッパー エンド ポイント値 - ノーマル オペレーティング コンディションでのアナログ出力の上限値のことです。 | R | Float | 64 |
vol_flow_analog_lower_limit | 469E | アナログ Lower リミット | アナログ アウトプット ロウアー エンド ポイント値 - ノーマル オペレーティング コンディションでのアナログ出力の下限値のことです。 | R | Float | 64 |
vol_flow_additional_analog_damping | 469F | Added Damping | アナログ出力追加ダンピング - ダンピング タイムはアナログ出力に作用します。単位は常に秒で表します。 | R/W | Float | 32 |
vol_flow_percent_range | 46A0 | Volume Flow Percent Range | % レンジ - ノーマル オペレーティング モードでの Lower/Upper レンジ値によるデジタル値表示です。単位は常にパーセントです。 | R | Float | 34 |
vol_flow_upper_range_value | 4606 | Volume Flow Upper Range Value | URV (アッパー レンジ バリュー) - アナログ出力に対するオペレーション上のエンド ポイントのことで, パーセント レンジの 100% に相当します。 | R/W | Float | 32 |
vol_flow_lower_range_value | 4607 | Volume Flow Lower Range Value | LRV (ロウアー レンジ バリュー) - アナログ出力に対するオペレーション上のエンド ポイントのことで, パーセント レンジの 0% に相当します。 | R/W | Float | 32 |