Name ID Label Description Type Access Class
fo_fault_level 43A1 Fault Level This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. R/W Float 64
density_for_cal 406B R/W Float 4096
point_code 406C R/W Unsigned 4096
densityfd_for_cal 406E R/W Float 4096
temp_for_cal 4373 Ref temperature User-entered temperature for the purpose of performing an on-line temperature calibration. R/W Float 4096
D1 4089 D1 The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. R/W Float 8
D2 408B D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. R/W Float 8
D3 4073 D3 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. R/W Float 8
D4 4075 D4 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. R/W Float 8
K1 408A K1 The tube period, in microseconds, corresponding to a density equal to D1. R/W Float 8