Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
cust_trans_totalizer_reset | 401D | Totalizer Reset Option | Custody transfer totalizer reset options | R/W | Enum | 32 |
World_area_weights_measures_security | 40F1 | Custody Transfer Approval | Displays World Area for Weights and Measures Security | R | Enum | 2048 |
dens_tgco | 4047 | DTG | This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco1 | 4048 | DFQ1 | This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco2 | 4049 | DFQ2 | This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_tgco | 404A | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_fqco | 404B | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | R/W | Float | 8 |
max_zero | 404E | Zero Time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | R/W | Unsigned | 12 |
pulse_width | 4051 | Maximum Pulse Width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | R/W | Float | 512 |
tempcal | 405C | Temperature Calibration Factor | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | R/W | ASCII | 32 |
Name | ID | Label | Description | Type | Access | Class |