Loading…
Name ID Label Description Type Access Class
zero_pres_factor 4094 Flow Calibration Pressure Pressure at which flow calibration was performed. R/W Float 8
enable_pres_comp 4095 Enable Pressure Compensation This parameter allows you to enable/disable the pressure correction feature R/W Enum 8
static_pressure 4096 Static Pressure This parameter allows you to enter a static pressure when using the pressure correction feature R/W Float 8
enable_extern_temp 40C9 Enable External Temperature This parameter allows you to enable/disable the external temperature feature R/W Enum 8
static_temperature 40CA Static Temperature This parameter allows you to enter a static temperature when using the external temperature feature R/W Float 8
flow_switch_setpoint 402D Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
flow_switch_variable_assignment 402F Source R/W Enum 8
do1_assignment 4570 Assignment This parameter identifies which condition the discrete output will act upon. R/W Enum 1024
do1_state 43E9 Discrete Output State This parameter identifies the state of the discrete output. R/W Enum 2
do1_polarity 4571 Polarity This parameter allows you to set the polarity of the discrete output. R/W Enum 1024
Name ID Label Description Type Access Class