Name ID Label Description Type Access Class
ao_1_fault_code 4141 Fault Action Downscale: mA outputs that indicate flow go to the configured limit,the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
ao_2_fault_code 4145 Fault Action Downscale: mA outputs that indicate flow go to the configured limit,the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
fo_fault_code 4147 Fault Action Downscale: mA outputs that indicate flow go to the configured limit,the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
dig_comm_fault_code 4149 Fault Action This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
ao_1_fault_level 4143 Fault Level This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. R/W Float 64
ao_2_fault_level 4146 Fault Level This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. R/W Float 64
fo_fault_level 4148 Fault Level This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. R/W Float 64
ma_1_cutoff 402D mA Output Flow Rate Cutoff The mass or volume flow rate below which is a mA output representing flow indicates zero flow. R/W Float 8
ma_2_cutoff 402F mA Output Flow Rate Cutoff The mass or volume flow rate below which is a mA output representing flow indicates zero flow. R/W Float 8
cutoff_code 414E Cutoff Type R/W Index 4096