Loading…
Name ID Label Description Type Access Class
drive_upper_range_value_1 409D Upper Range Value Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
drive_lower_range_value_1 409E Lower Range Value Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
drive_upper_sensor_limit 409F Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
drive_lower_sensor_limit 40A0 Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
drive_minimum_span 40A1 Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
api_dens_digital_value 40A4 Density at Reference Temperature R Float 10
api_dens_analog_value 40A5 Density at Reference Temperature R/W Float 66
api_dens_analog_value_1 40A6 Density at Reference Temperature R/W Float 66
api_dens_digital_cutoff 4366 Density Cutoff The density below which the density indicates zero. R/W Float 8
api_dens_digital_damping 4367 Density Damping Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective mA output. The units of this variable are always in seconds. R/W Float 32
Name ID Label Description Type Access Class