名称 ID 标签 描述 类型 访问 分类
method_flag 489C Set Frequency Output Scaling 读/写 枚举 4096
method_flag_ip 48A3 设置频率输入定标 读/写 枚举 4096
method_float_placeholder 4AFB 读/写 浮点 4096
core_firmware_checksum 4654 Core Processor 无符号 2048
transmitter_firmware_checksum 4655 Transmitter 无符号 2048
weights_measures_software_version 4656 重量和测试软件版本 无符号 0
update_rate 4667 Update Rate This parameter is the update rate for process variables through the system. Set the update rate to NORMAL for most applications. Set the update rate to SPECIAL if you are using the device in proving applications or extremely short batch applications. Refer to the user's manual for more information before you change this parameter. 读/写 枚举 2048
update_rate_var 4668 100 Hz Variable This parameter identifies the transmitter variable that will be updated at 100 Hz. All others will be updated, but at a slower rate. 读/写 枚举 2048
varGas_Density_Unit 465E Gas Density Unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. 读/写 枚举 32
flow_switch_hysteresis 4653 Hysteresis (0.1-10.0) If the hysteresis is set to 5%, no change of the Discrete Output state occurs when the flow rate is within +-5% of the setpoint. If the setpoint is 100 lb/min and the first reading is below 95 lb/min, the flow switch turns On and will stay On until the flow rate rises above 105 lb/min. At this point it turns Off, and will stay off until the flow rate drops below 95 lb/min. 读/写 浮点 8