Name ID Label Description Type Access Class
tempcal 414F Temperature Calibration Factor A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. R/W ASCII 32
temp_for_cal 4150 Контрольная температура User-entered temperature for the purpose of performing an on-line temperature calibration. R/W Float 4096
density_for_cal 4151 R/W Float 4096
point_code 4152 R/W Unsigned 4096
densityfd_for_cal 4153 R/W Float 4096
D1 4154 D1 The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. R/W Float 8
D2 4155 D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. R/W Float 8
D3 4156 D3 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. R/W Float 8
D4 4157 D4 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. R/W Float 8
K1 4158 K1 The tube period, in microseconds, corresponding to a density equal to D1. R/W Float 8
Name ID Label Description Type Access Class