Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
fi_flow_rate_factor | 4132 | Rate Factor | The flow rate that equals a set frequency if the frequency input represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency input represents mass or volume total. | R/W | Float | 512 |
fi_pulsesPerUnit | 4133 | Pulses/Unit | This parameter allows you to scale the frequency input in terms of the number of pulses per unit of flow. | R/W | Float | 512 |
fi_unitsPerPulse | 4134 | Units/Pulse | This parameter allows you to scale the frequency input in terms of the number of units of flow required to generate one pulse. | R/W | Float | 512 |
fi_k_factor | 4135 | K Factor | The number of frequency pulses per unit of flow that the frequency is inputting. For example 10 pulses/kg. | R/W | Float | 512 |
max_zero | 4136 | Zero Time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | R/W | Unsigned | 12 |
rt_zero | 4137 | Zero Value | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | R/W | Float | 10 |
zero_offset_units | 4138 | Zero units | Units for zero offset - always microseconds. | R/W | Unsigned | 32 |
rt_sigma | 4139 | Standard Deviation | A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. | R | Float | 10 |
fault_index | 413A | Output Type | R/W | Index | 4096 | |
LMV_fault_timeout | 413C | Время задержки в случае неисправности | This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. | R/W | Unsigned | 12 |
Name | ID | Label | Description | Type | Access | Class |