Loading…
Name ID Label Description Type Access Class
gas_vol_flow_analog_value_1 44EF Vol flo anlog valu R/W Float 66
gas_vol_flow_additional_analog_damping_1 44F0 Added damp Analog Output Added Damping Value- Damping time constant applied the Analog Output. The units of this variable are always in seconds. R/W Float 32
gas_vol_flow_percent_range_1 44F1 Vol flo % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
gas_vol_flow_upper_range_value_1 44F2 Vol flo URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
gas_vol_flow_lower_range_value_1 44F3 Vol flo LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
gas_vol_flow_upper_sensor_limit 44F4 Vol flo URL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
gas_vol_flow_lower_sensor_limit 44F5 Vol flo LRL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
gas_vol_flow_minimum_span 44F6 Vol flo min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
gas_vol_flow_hzfreq 44F7 Freq factr The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. R/W Float 512
gas_vol_flow_hzrate 44F8 Rate factr The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512
Name ID Label Description Type Access Class