Loading…
名称 ID 标签 描述 类型 访问 分类
sluglo 411A Slug low limit A programmable low limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. 读/写 浮点 8
slug_duration 411B Slug duration A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. 读/写 浮点 12
flocal 411C FlowCal A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. 读/写 ASCII 32
fcf 411D FCF A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. 读/写 ASCII 32
sp_mass 411E Base mass unit A standard engineering unit that serves as the mass component in a special unit. 读/写 枚举 32
sp_mass_time 411F Base mass time A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. 读/写 枚举 32
sp_mass_fact 4120 Mass flo conv factr A programmable divider for the base mass unit in a special unit. 读/写 浮点 32
sp_mass_rate_text 4121 Mass flo text A 1 to 8 character description of a special unit of flow. 读/写 ASCII 2048
sp_mass_total_text 4122 Mass totl text A 1 to 8 character description of a special unit of mass, which is based on the special unit of flow to represent total. 读/写 ASCII 2048
sp_volume 4123 Base vol unit A standard engineering unit that serves as the volume component in a special unit. 读/写 枚举 32
名称 ID 标签 描述 类型 访问 分类