Loading…
Name ID Label Description Type Access Class
static_pressure 41DB Static Pressure This parameter allows you to enter a static pressure when using the pressure correction feature R/W Float 8
enable_extern_temp 41DC Enable Ext Temp This parameter allows you to enable/disable the external temperature feature R/W Enum 8
static_temperature 41DD Static Temperature This parameter allows you to enter a static temperature when using the external temperature feature R/W Float 8
flow_switch_setpoint 41DE Flow Switch Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
do1_assignment 41DF DO 1 is This parameter identifies which condition the discrete output will act upon. R/W Enum 1024
do1_state 41E0 DO is This parameter identifies the state of the discrete output. R/W Enum 1024
do1_polarity 41E1 DO 1 Polarity This parameter allows you to set the polarity of the discrete output. R/W Enum 1024
do1_FaultIndication 4389 DO 1 Fault Indication R/W Enum 1024
do2_assignment 41E2 DO 2 is This parameter identifies which condition the discrete output will act upon. R/W Enum 1024
do2_state 41E3 DO is This parameter identifies the state of the discrete output. R/W Enum 1024
Name ID Label Description Type Access Class