Loading…
Name ID Label Description Type Access Class
D2 4155 D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. R/W Float 8
D3 4156 D3 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. R/W Float 8
D4 4157 D4 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. R/W Float 8
K1 4158 K1 The tube period, in microseconds, corresponding to a density equal to D1. R/W Float 8
K2 4159 K2 The tube period, in microseconds, corresponding to a density equal to D2. R/W Float 8
K3 415A K3 The tube period, in microseconds, corresponding to a density equal to D3. R/W Float 8
K4 415B K4 The tube period, in microseconds, corresponding to a density equal to D4. R/W Float 8
FD 415C FD A calibration constant generated during a flowing density calibration. R/W Float 8
DCoef 415D Temp coeff The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8
DT 415E DT The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8
Name ID Label Description Type Access Class