Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
fi_input_freq_digital_units | 4650 | FI Freq | The frequency, in Hertz, presently being indicated on the frequency input. | R | Enum | 32 |
fi_input_frequency | 412E | FI Freq | R | Unsigned | 514 | |
fi_scaling_method | 412F | FI Scale Method | This parameter allows you to scale the frequency input by using freq=flow, pulses/unit, or units/pulse. This parameter must be sent to the device before you make scaling changes. | R/W | Enum | 512 |
fi_freq_factor | 4131 | Freq factr | The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency input. | R/W | Float | 512 |
fi_flow_rate_factor | 4132 | Rate factr | The flow rate that equals a set frequency if the frequency input represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency input represents mass or volume total. | R/W | Float | 512 |
fi_pulsesPerUnit | 4133 | Pulses/Unit | This parameter allows you to scale the frequency input in terms of the number of pulses per unit of flow. | R/W | Float | 512 |
fi_unitsPerPulse | 4134 | Units/Pulse | This parameter allows you to scale the frequency input in terms of the number of units of flow required to generate one pulse. | R/W | Float | 512 |
fi_k_factor | 4135 | K Factor | The number of frequency pulses per unit of flow that the frequency is inputting. For example 10 pulses/kg. | R/W | Float | 512 |
max_zero | 4136 | Zero time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | R/W | Unsigned | 12 |
rt_zero | 4137 | Zero | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | R/W | Float | 10 |
Name | ID | Label | Description | Type | Access | Class |