Name ID Label Description Type Access Class
static_temperature 4211 Температура корректировки This parameter allows you to enter a static temperature when using the external temperature feature R/W Float 8
flow_switch_setpoint 4128 Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
do_index 412A DO index R/W Index 4096
flow_switch_hysteresis 4588 Hysteresis (0.1-10.0) If the hysteresis is set to 5%, no change of the Discrete Output state occurs when the flow rate is within +-5% of the setpoint. If the setpoint is 100 lb/min and the first reading is below 95 lb/min, the flow switch turns On and will stay On until the flow rate rises above 105 lb/min. At this point it turns Off, and will stay off until the flow rate drops below 95 lb/min. R/W Float 8
chan_a 412E Назначение This parameter identifies the physical output channel to be configured. Channel A is terminals 1 and 2, Channel B is terminals 3 and 4, and Channel C is terminals 5 and 6. R Enum 64
chan_b 4006 Назначение This parameter identifies the physical output channel to be configured. Channel A is terminals 1 and 2, Channel B is terminals 3 and 4, and Channel C is terminals 5 and 6. R/W Enum 64
chan_c 412F Назначение This parameter identifies the physical output channel to be configured. Channel A is terminals 1 and 2, Channel B is terminals 3 and 4, and Channel C is terminals 5 and 6. R Enum 64
current_ed_curve 4238 Активная матрица The Enhanced Density Curve that is currently selected. R/W Enum 2048
ed_std_vol_flow_digital_value 4239 Flow Rate R Float 10
edstdvolflowClassificationCode 4D8F Классификация стандартного объёмного расхода по ED Классификация - указывает таблицу расширения кода единицы для конкретной динамической переменной или переменной устройства R Enum 0
Name ID Label Description Type Access Class