名称 ID 标签 描述 类型 访问 分类
tempcal 40BF Temperature Calibration Factor A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. 读/写 ASCII 32
rt_zero 40C0 Zero Value A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. 读/写 浮点 10
zero_offset_units 40C1 Zero units Units for zero offset - always microseconds. 读/写 无符号 32
rt_sigma 40C2 Standard Deviation A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. 浮点 10
fault_index 40C3 Output Type 读/写 索引 4096
LMV_fault_timeout 40C5 Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. 读/写 无符号 12
ao_1_fault_code 40CA 故障动作 Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. 读/写 枚举 12
ao_2_fault_code 40CE 故障动作 Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. 读/写 枚举 12
fo_fault_code 40D0 故障动作 Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. 读/写 枚举 12
dig_comm_fault_code 40D2 故障动作 This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) 读/写 枚举 12
名称 ID 标签 描述 类型 访问 分类