Name ID Label Description Type Access Class
update_rate_var 4124 100 Hz Variable This parameter identifies the transmitter variable that will be updated at 100 Hz. All others will be updated, but at a slower rate. R/W Enum 2048
flow_pres_factor 4125 Flow Pressure Factor Pressure compensation factor applied to flow rate (% change per PSI). R/W Float 8
dens_pres_factor 4126 Density Pressure Factor Pressure compensation factor applied to density (g/Cucm change per PSI). R/W Float 8
zero_pres_factor 4127 Flow Calibration Pressure Pressure at which flow calibration was performed. R/W Float 8
enable_pres_comp 41FC Pressure Compensation This parameter allows you to enable/disable the pressure correction feature R/W Enum 8
static_pressure 41FD Compensation Pressure This parameter allows you to enter a static pressure when using the pressure correction feature R/W Float 8
enable_extern_temp 4210 External Temperature This parameter allows you to enable/disable the external temperature feature R/W Enum 8
static_temperature 4211 Correction Temperature This parameter allows you to enter a static temperature when using the external temperature feature R/W Float 8
flow_switch_setpoint 4128 Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
do_index 412A DO index R/W Index 4096