Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
live_zero_vol_flow_hzfreq | 4428 | Frequency Factor | The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. | R/W | Float | 512 |
live_zero_vol_flow_hzrate | 4429 | Rate Factor | The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. | R/W | Float | 512 |
live_zero_vol_flow_p_u | 442A | Pulses/Unit | This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. | R/W | Float | 512 |
live_zero_vol_flow_u_p | 442B | Units/Pulse | This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. | R/W | Float | 512 |
field_ver_zero_digital_units | 4590 | Field Verification Zero | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. | R/W | Enum | 32 |
field_ver_zero_digital_value | 4591 | Field Verification Zero | R | Float | 10 | |
cust_trans_totalizer_reset | 442C | Totalizer Reset Option | Custody transfer totalizer reset options | R/W | Enum | 32 |
factory_zero_offset | 442D | Factory zero offset | Factory zero offset value - always microseconds. | R/W | Float | 10 |
discrete_action_code | 442E | Discrete Action | Selects an action to be assigned to a discrete. | R/W | Enum | 4096 |
discrete_assignment_code | 442F | Discrete Assignment | Assigns a discrete to an action. Select none for no discrete assignment. | R/W | Enum | 4096 |
Name | ID | Label | Description | Type | Access | Class |