Loading…
名称 ID 标签 描述 类型 访问 分类
sluglo 40AB Slug low limit A programmable low limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. 读/写 浮点 8
slug_duration 40AC Slug duration A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. 读/写 浮点 12
flocal 40AD FlowCal A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. 读/写 ASCII 32
fcf 40AF FCF A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. 读/写 ASCII 32
ma_1_cutoff 4028 PV AO cutoff The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. 读/写 浮点 8
ma_2_cutoff 402A SV AO cutoff The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. 读/写 浮点 8
sp_volume 40B0 Base vol unit A standard engineering unit that serves as the volume component in a special unit. 读/写 枚举 32
sp_volume_time 40B1 Base vol time A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. 读/写 枚举 32
sp_volume_fact 40B2 Vol flo conv factr A programmable divider for the base volume unit in a special unit. 读/写 浮点 32
sp_volume_rate_text 40B3 Vol flo text A 1 to 8 character description of a special unit of flow. 读/写 ASCII 2048
名称 ID 标签 描述 类型 访问 分类