Name ID Label Description Type Access Class
dens_factor 40FB Dens factor A value, between 0.8 and 1.2, that will act as a multiplier for the density. R/W Float 8
tube_freq 40FC Tube The tube frequency of the sensor measured in Hertz. R Float 2
tube_amplitude_L 40FD LPO The pickoff voltage of the sensor measured in Voltage. R Float 2
tube_amplitude_R 40FE RPO The pickoff voltage of the sensor measured in Voltage. R Float 2
drive_gain 40FF Drive The drive gain of the transmitter. R Float 2
live_zero 4100 Live zero flow This flow value is not effected by the flow cutoff and is damped at 8 seconds. It is to be used for diagnostic purposes only and is not used in ANY flow calculations or totalization. R Float 2
board_temp_700 4101 Board Temperature The board temperature of the core processor. R Float 2
case_temp 4102 Meter Temp: T-Series The meter temperature of the T-Series sensor. R Float 2
FD 4103 FD A calibration constant generated during a flowing density calibration. R/W Float 8
tot_reset 4104 Disp. Total Reset When this parameter is enabled, you can use the display to reset the totalizer. R/W Enum 256
Name ID Label Description Type Access Class