Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
totctl | 40B6 | Управление сумматором | Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. | R/W | Enum | 4096 |
fo_scaling_method | 4022 | FO Scale Method | This parameter allows you to scale the frequency output by using freq=flow, pulses/unit, or units/pulse. This parameter must be sent to the device before you make scaling changes. | R/W | Enum | 512 |
dens_tgco | 40B8 | DTG | This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco1 | 40B9 | DFQ1 | This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco2 | 40BA | DFQ2 | This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_tgco | 40BB | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_fqco | 40BC | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | R/W | Float | 8 |
max_zero | 40BD | Zero time | The number of readings collected from the sensor during an Auto Zero operation. | R/W | Unsigned | 4 |
pulse_width | 40BE | Max pulse width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | R/W | Float | 512 |
tempcal | 40BF | Temp cal factr | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | R/W | ASCII | 32 |