Name ID Label Description Type Access Class
poll_type_code 411D Polling Type This parameter identifies which compensation algorithms will be enabled within the device based on the polling being done. R/W Enum 128
update_rate 4123 Update Rate This parameter is the update rate for process variables through the system. Set the update rate to NORMAL for most applications. Set the update rate to SPECIAL if you are using the device in proving applications or extremely short batch applications. Refer to the user's manual for more information before you change this parameter. R/W Enum 2048
update_rate_var 4124 Update rate var This parameter identifies the transmitter variable that will be updated at 100 Hz. All others will be updated, but at a slower rate. R/W Enum 2048
flow_pres_factor 4125 Flow factr Pressure compensation factor applied to flow rate (% change per PSI). R/W Float 8
dens_pres_factor 4126 Dens factr Pressure compensation factor applied to density (g/Cucm change per PSI). R/W Float 8
zero_pres_factor 4127 Flowcal pressure Pressure at which flow calibration was performed. R/W Float 8
flow_switch_setpoint 4128 Flow Switch Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
do1_assignment 4129 DO is This parameter identifies which condition the discrete output will act upon. R/W Enum 1024
do_index 412A DO index R/W Index 4096
chan_a 412E Channel A This parameter identifies the physical output channel to be configured. Channel 1 is terminals 1 & 2, Channel 2 is terminals 3 & 4 and Channel 3 is terminals 5 & 6. R Enum 64