Name ID Label Description Type Access Class
rt_zero 40C0 Zero A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. R/W Float 2
zero_offset_units 40C1 Zero units Units for zero offset - always microseconds. R/W Unsigned 32
rt_sigma 40C2 Std dev A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 2
fault_index 40C3 Output Type R/W Index 4096
LMV_fault_timeout 40C5 Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. R/W Unsigned 4
ao_1_fault_code 40CA AO1 Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 4
ao_2_fault_code 40CE AO2 Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 4
fo_fault_code 40D0 FO Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 4
dig_comm_fault_code 40D2 Comm Fault ind. This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 4
ao_1_fault_level 40CC mA1 Fault Value This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. R/W Float 64
Name ID Label Description Type Access Class