Name ID Label Description Type Access Class
mass_flow_hzrate 4023 Rate factr The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512
mass_flow_p_u 4024 Pulses/Unit This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. R/W Float 512
mass_flow_u_p 4025 Units/Pulse This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. R/W Float 512
temp_digital_units 4031 Temp unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. R/W Enum 32
temp_digital_value 4032 Temp R Float 2
temp_analog_value 4033 Temp anlog valu R/W Float 66
temp_analog_upper_limit 4034 Límite sup. analóg. Analog Output Upper Endpoint Value- Defines the upper endpoint, of the respective Analog Output, for normal operating conditions R Float 64
temp_analog_lower_limit 4035 Límite inf. analóg. Analog Output Lower Endpoint Value- Defines the lower endpoint, of the respective Analog Output, for normal operating conditions R Float 64
temp_additional_analog_damping 4036 Added damp Analog Output Added Damping Value- Damping time constant applied the Analog Output. The units of this variable are always in seconds. R/W Float 32
temp_percent_range 4037 Temp % rnge Porcentaje de rango: Variable que rastrea la representación del valor digital con respecto al rango definido por el valor de rango inferior y el valor de rango superior, para modos de funcionamiento normales. Las unidades de esta variable siempre son en porcentaje. R Float 34
Name ID Label Description Type Access Class