Name ID Label Description Type Access Class
dens_minimum_span 405B Dens min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
minv_digital_value 405E Mass inventory R Float 2
vol_flow_digital_units 4060 Vol flo unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. R/W Enum 32
vol_flow_digital_value 4061 Vol flo R Float 2
vol_flow_digital_cutoff 4063 Vol flo cutoff The volume flow rate below which the frequency/pulse output indicates zero flow, and below which totalizers stop counting. R/W Float 8
vol_flow_analog_value 4064 Vol flo anlog valu R/W Float 66
vol_flow_analog_upper_limit 4065 Anlog upr limit Analog Output Upper Endpoint Value- Defines the upper endpoint, of the respective Analog Output, for normal operating conditions R Float 64
vol_flow_analog_lower_limit 4066 Anlog lor limit Analog Output Lower Endpoint Value- Defines the lower endpoint, of the respective Analog Output, for normal operating conditions R Float 64
vol_flow_additional_analog_damping 4067 Added damp Analog Output Added Damping Value- Damping time constant applied the Analog Output. The units of this variable are always in seconds. R/W Float 32
vol_flow_percent_range 4068 Vol flo % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34